Instructions
RIGOL Chapter 2 Command System
2-108 DS1000Z-E Programming Guide
:MEASure:SETup:DSB
Syntax :MEASure:SETup:DSB <source>
:MEASure:SETup:DSB?
Description
Set or query source B of Delay
1→2 and Delay 1→2 measurements.
Parameter
Name
Type
Range
Default
<source>
Discrete
{CHANnel1|CHANnel2}
CHANnel2
Return
Format
The query returns CHAN1 or CHAN2.
Example
:MEASure:SETup:DSB CHANnel2 /*Set source B of delay measurement to CH2*/
:MEASure:SETup:DSB? /*The query returns CHAN2*/
Related
Command
:MEASure:ITEM
:MEASure:STATistic:DISPlay
Syntax :MEASure:STATistic:DISPlay <bool>
:MEASure:STATistic:DISPlay?
Description
Enable or disable the statistic function, or query the status of the statistic function.
Parameter
Name
Type
Range
Default
<bool>
Bool
{{1|ON}|{0|OFF}}
0|OFF
Explanation When the statistic function is enabled, the oscilloscope makes statistic and displays the
statistic results of at most 5 measurement items that are turned on last.
Return
Format
The query returns 1 or 0.
Example
:MEASure:STATistic:DISPlay ON /*Enable the statistic function*/
:MEASure:STATistic:DISPlay? /*The query returns 1*/
Related
Commands
:MEASure:STATistic:MODE
:MEASure:STATistic:RESet
:MEASure:STATistic:ITEM
:MEASure:ITEM
:MEASure:STATistic:MODE
Syntax :MEASure:STATistic:MODE <mode>
:MEASure:STATistic:MODE?
Description
Set or query the statistic mode.
Parameter
Name
Type
Range
Default
<mode>
Discrete
{DIFFerence|EXTRemum}
EXTRemum
Explanation DIFFerence: select the difference mode. The statistic results contain the current
value, average value, standard deviation, and counts.
EXTRemum: select the extremum mode. The statistic results contain the current
value, average value, minimum, and maximum.