Instructions
Chapter 2 Command System RIGOL
DS1000Z-E Programming Guide 2-109
Sending the :MEASure:STATistic:DISPlay command can enable the statistic function.
When the statistic function is enabled, the oscilloscope makes statistic and displays
the statistic results of at most 5 measurement items that are turned on last.
Return
Format
The query returns DIFF or EXTR.
Example
:MEASure:STATistic:MODE DIFFerence /*Select the difference mode*/
:MEASure:STATistic:MODE? /*The query returns DIFF*/
Related
Commands
:MEASure:STATistic:RESet
:MEASure:STATistic:ITEM
:MEASure:STATistic:RESet
Syntax
:MEASure:STATistic:RESet
Description
Clear the history data and make statistic again.
Related
Commands
:MEASure:STATistic:DISPlay
:MEASure:STATistic:MODE
:MEASure:STATistic:ITEM
:MEASure:STATistic:ITEM
Syntax :MEASure:STATistic:ITEM <item>[,<src>[,<src>]]
:MEASure:STATistic:ITEM? <type>,<item>[,<src>[,<src>]]
Description Enable the statistic function of any waveform parameter of the specified source, or query
the statistic result of any waveform parameter of the specified source.
Parameter
Name Type Range Default
<item> Discrete
{VMAX|VMIN|VPP|VTOP|VBASe|VAMP|VAVG|VRMS|
OVERshoot|PREShoot|MARea|MPARea|PERiod|FRE
Quency|RTIMe|FTIMe|PWIDth|NWIDth|PDUTy|ND
UTy|RDELay|FDELay|RPHase|FPHase|TVMAX|TVMI
N|PSLEWrate|NSLEWrate|VUPper|VMID|VLOWer|VA
RIance|PVRMS|PPULses|NPULses|PEDGes|NEDGes}
--
<src> Discrete {CHANnel1|CHANnel2|MATH}
<type> Discrete
{MAXimum|MINimum|CURRent|AVERages|
DEViation}
--
Explanation [,<src>[,<src>]] sets the source of the parameter to be measured.
If the parameter to be measured only needs a single source (VMAX, VMIN, VPP,
VTOP, VBASe, VAMP, VAVG, VRMS, OVERshoot, MARea, MPARea, PREShoot, PERiod,
FREQuency, RTIMe, FTIMe, PWIDth, NWIDth, PDUTy, NDUTy, TVMAX, TVMIN,
PSLEWrate, NSLEWrate, VUPper, VMID, VLOWer, VARIance, PVRMS, PPULses,
NPULses, PEDGes, and NEDGes), you only need to set a single source. If
[,<src>[,<src>]] is omitted, the source is the one selected by the
:MEASure:SOURce
command by default.
If the parameter to be measured needs two sources (RDELay, FDELay, RPHase, and
FPHase), the command needs to include two sources; otherwise, the command is
invalid. If [,<src>[,<src>]] is omitted, the sources are the ones selected by
the :MEASure:SETup:DSA and :MEASure:SETup:DSB or :MEASure:SETup:PSA