Instructions

RIGOL Chapter 2 Command System
2-134 DS1000Z-E Programming Guide
Format
Example :TRIGger:SLOPe:WHEN NLESs /*Set the trigger condition to NLESs*/
:TRIGger:SLOPe:WHEN? /*The query returns NLES*/
:TRIGger:SLOPe:TIME
Syntax :TRIGger:SLOPe:TIME <time>
:TRIGger:SLOPe:TIME?
Description
Set or query the time value in slope trigger. The default unit is s.
Parameter
Name
Type
Range
Default
<time> Real 8ns to 10s
PGReater, NGReater: 1μs
PLESs, NLESs: 2μs
Explanation This command is available when the trigger condition (refer to :TRIGger:SLOPe:WHEN) is
PGReater, PLESs, NGReater, and NLESs.
Return
Format
The query returns the time value in scientific notation.
Example
:TRIGger:SLOPe:TIME 0.000003 /*Set the time value to 3μs*/
:TRIGger:SLOPe:TIME? /*The query returns 3.000000e-06*/
:TRIGger:SLOPe:TUPPer
Syntax :TRIGger:SLOPe:TUPPer <time>
:TRIGger:SLOPe:TUPPer?
Description
Set or query the upper limit of the time in slope trigger. The default unit is s.
Parameter
Name
Type
Range
Default
<time>
Real
16ns to 10s
2μs
Explanation This command is available when the trigger condition (refer to :TRIGger:SLOPe:WHEN) is
PGLess and NGLess.
Return
Format
The query returns the upper limit of the time in scientific notation.
Example
:TRIGger:SLOPe:TUPPer 0.000003 /*Set the upper limit of the time to 3μs*/
:TRIGger:SLOPe:TUPPer? /*The query returns 3.000000e-06*/
Related
Command
:TRIGger:SLOPe:TLOWer
:TRIGger:SLOPe:TLOWer
Syntax :TRIGger:SLOPe:TLOWer <time>
:TRIGger:SLOPe:TLOWer?
Description
Set or query the lower limit of the time in slope trigger. The default unit is s.
Parameter
Name
Type
Range
Default
<time>
Real
8ns to 9.99s
1μs
Explanation This command is available when the trigger condition (refer to :TRIGger:SLOPe:WHEN) is
PGLess and NGLess.