Instructions
Chapter 2 Command System RIGOL
DS1000Z-E Programming Guide 2-145
:TRIGger:TIMeout
Command List:
:TRIGger:TIMeout:SOURce
:TRIGger:TIMeout:SLOPe
:TRIGger:TIMeout:TIMe
:TRIGger:TIMeout:SOURce
Syntax :TRIGger:TIMeout:SOURce <source>
:TRIGger:TIMeout:SOURce?
Description
Set or query the trigger source in timeout trigger.
Parameter
Name
Type
Range
Default
<source>
Discrete
{CHANnel1|CHANnel2}
CHANnel1
Return
Format
The query returns CHAN1 or CHAN2.
Example
:TRIGger:TIMeout:SOURce CHANnel2 /*Set the trigger source to CH2*/
:TRIGger:TIMeout:SOURce? /*The query returns CHAN2*/
:TRIGger:TIMeout:SLOPe
Syntax :TRIGger:TIMeout:SLOPe <slope>
:TRIGger:TIMeout:SLOPe?
Description
Set or query the edge type in timeout trigger.
Parameter
Name
Type
Range
Default
<slope>
Discrete
{POSitive|NEGative|RFALl}
POSitive
Explanation POSitive: start timing when the rising edge of the input signal passes through the
trigger level.
NEGative: start timing when the falling edge of the input signal passes through the
trigger level.
RFALl: start timing when any edge of the input signal passes through the trigger
level.
Return
Format
The query returns POS, NEG, or RFAL.
Example
:TRIGger:TIMeout:SLOPe NEGative /*Set the edge type to falling edge*/
:TRIGger:TIMeout:SLOPe? /*The query returns NEG*/
:TRIGger:TIMeout:TIMe
Syntax :TRIGger:TIMeout:TIMe <NR3>
:TRIGger:TIMeout:TIMe?
Description
Set or query the timeout time in timeout trigger. The default unit is s.
Parameter
Name
Type
Range
Default
<NR3>
Real
16ns to 10s
16ns