Instructions
Chapter 2 Command System RIGOL
DS1000Z-E Programming Guide 2-177
Command List:
:WAVeform:SOURce
:WAVeform:MODE
:WAVeform:FORMat
:WAVeform:DATA?
:WAVeform:XINCrement?
:WAVeform:XORigin?
:WAVeform:XREFerence?
:WAVeform:YINCrement?
:WAVeform:YORigin?
:WAVeform:YREFerence?
:WAVeform:STARt
:WAVeform:STOP
:WAVeform:PREamble?
:WAVeform:SOURce
Syntax :WAVeform:SOURce <source>
:WAVeform:SOURce?
Description
Set or query the channel of which the waveform data will be read.
Parameter
Name
Type
Range
Default
<source> Discrete {CHANnel1|CHANnel2|MATH} CHANnel1
Explanation
If the MATH channel is selected, only NORMal can be selected in :WAVeform:MODE.
Return
Format
The query returns CHAN1, CHAN2, or MATH.
Example :WAVeform:SOURce CHANnel2 /*Set the channel to CH2*/
:WAVeform:SOURce? /*The query returns CHAN2*/
Related
Command
:WAVeform:FORMat
:WAVeform:MODE
Syntax :WAVeform:MODE <mode>
:WAVeform:MODE?
Description Set or query the reading mode used by :WAVeform:DATA?.
Parameter
Name
Type
Range
Default
<mode> Discrete {NORMal|MAXimum|RAW} NORMal
Explanation NORMal: read the waveform data displayed on the screen.
MAXimum: read the waveform data displayed on the screen when the instrument is
in the run state and the waveform data in the internal memory in the stop state.