Instructions
Chapter 2 Command System RIGOL
DS1000Z-E Programming Guide 2-183
:WAVeform:YORigin?
Syntax
:WAVeform:YORigin?
Description Query the vertical offset relative to the vertical reference position of the specified channel
source in the Y direction.
Explanation The return value is related to the current data reading mode:
In NORMal mode, YORigin = VerticalOffset/YINCrement.
In RAW mode, YORigin is related to the Verticalscale of the internal waveform and the
Verticalscale currently selected.
In MAX mode, YORigin = VerticalOffset/YINCrement when the instrument is in running
status; YORigin is related to the Verticalscale of the internal waveform and the
Verticalscale currently selected when the instrument is in stop status.
Return
Format
The query returns an integer.
Example
:WAVeform:YORigin? /*The query returns 25*/
Related
Commands
:WAVeform:SOURce
:WAVeform:YREFerence?
:WAVeform:YREFerence?
Syntax
:WAVeform:YREFerence?
Description
Query the vertical reference position of the specified channel source in the Y direction.
Explanation
YREFerence is always 127 (the screen bottom is 0 and the screen top is 255).
Return
Format
The query returns 127.
Related
Command
:WAVeform:SOURce
:WAVeform:STARt
Syntax :WAVeform:STARt <sta>
:WAVeform:STARt?
Description
Set or query the start point of waveform data reading.
Parameter
Name
Type
Range
Default
<sta> Integer
NORMal: 1 to 1200
MAX: 1 to the number of effective points
currently on the screen
RAW: 1 to the current maximum memory depth
1
Explanation When reading the waveform data in the internal memory, the actual settable ranges of the
start point and stop point of a reading operation are related to the memory depth of the
oscilloscope and the return format of the waveform data currently selected; for the
details, refer to the explanation in the :WAVeform:DATA? command.
Return
Format
The query returns an integer.
Example
:WAVeform:STARt 100 /*Set the start point to 100*/
:WAVeform:STARt? /*The query returns 100*/
Related
:ACQuire:MDEPth