Instructions

Chapter 2 Command System RIGOL
DS1000Z-E Programming Guide 2-21
:CURSor:MANual:YDELta?
Syntax
:CURSor:MANual:YDELta?
Description Query the difference between the Y values of cursor A and cursor B (BY-AY) in the manual
cursor measurement mode. The unit depends on the vertical unit currently selected.
Return
Format
When the signal source is CHANnel1|CHANnel2|MATH, the query returns the difference in
scientific notation.
Example
:CURSor:MANual:YDELta? /*The query returns -4.000000e+00*/
Related
Commands
:CURSor:MANual:AY
:CURSor:MANual:BY
:CURSor:MANual:VUNit
:CURSor:TRACk
Command List:
:CURSor:TRACk:SOURce1
:CURSor:TRACk:SOURce2
:CURSor:TRACk:AX
:CURSor:TRACk:BX
:CURSor:TRACk:AY?
:CURSor:TRACk:BY?
:CURSor:TRACk:AXValue?
:CURSor:TRACk:AYValue?
:CURSor:TRACk:BXValue?
:CURSor:TRACk:BYValue?
:CURSor:TRACk:XDELta?
:CURSor:TRACk:YDELta?
:CURSor:TRACk:IXDELTA?
:CURSor:TRACk:SOURce1
Syntax :CURSor:TRACk:SOURce1 <source>
:CURSor:TRACk:SOURce1?
Description
Set or query the channel source of cursor A in the track cursor measurement mode.
Parameter
Name
Type
Range
Default
<source> Discrete {OFF|CHANnel1|CHANnel2|MATH} CHANnel1
Explanation
Only the channels enabled can be selected as the channel source.
Return
Format
The query returns OFF, CHAN1, CHAN2, or MATH.
Example
:CURSor:TRACk:SOURce1 CHANnel2 /*Set the channel source to CH2*/
:CURSor:TRACk:SOURce1? /*The query returns CHAN2*/