Instructions
RIGOL Chapter 2 Command System
2-82 DS1000Z-E Programming Guide
Example :MATH:SOURce2 CHANnel2 /*Set source B of algebraic operation to CH2*/
:MATH:SOURce2? /*The query returns CHAN2*/
:MATH:LSOUrce1
Syntax :MATH:LSOUrce1 <src>
:MATH:LSOUrce1?
Description
Set or query source A of logic operation.
Parameter
Name Type Range Default
<src> Discrete {CHANnel1|CHANnel2} CHANnel1
Explanation
The logic operations include A&&B, A||B, A^B, and !A.
Return
Format
The query returns CHAN1 or CHAN2.
Example
:MATH:LSOUrce1 CHANnel1 /*Set source A of logic operation to CH1*/
:MATH:LSOUrce1? /*The query returns CHAN1*/
:MATH:LSOUrce2
Syntax :MATH:LSOUrce2 <src>
:MATH:LSOUrce2?
Description Set or query source B of logic operation.
Parameter
Name
Type
Range
Default
<src> Discrete {CHANnel1|CHANnel2} CHANnel1
Explanation The logic operations include A&&B, A||B, A^B, and !A.
This command is only applicable to logic operations that require two signal sources
and is used to set source B.
Return
Format
The query returns CHAN1 or CHAN2.
Example
:MATH:LSOUrce2 CHANnel1 /*Set source B of logic operation to CH1*/
:MATH:LSOUrce2? /*The query returns CHAN1*/
:MATH:SCALe
Syntax :MATH:SCALe <scale>
:MATH:SCALe?
Description Set or query the vertical scale of the operation result. The unit depends on the operator
currently selected and the unit of the source.
Parameter
Name
Type
Range
Default
<scale>
Real
The max range is from 1p to 5T (in 1-2-5 step)
1.00V