Chapter 1 Quick Start RIGOL
DS1000Z-E User Guide 1-11
Pass/Fail:
The instrument can output a negative pulse via this connector when a failed
waveform is detected during the pass/fail test. The instrument continuously
outputs a low level via this connector when no failed waveform is detected.
5. Lock Hole
You can lock the instrument to a fixed location by using the security lock (please
purchase it yourself) via the lock hole.
6. Fuse
If a new fuse is required, please use the specified fuse (250V, T2A). The
replacing method is as follows.
a) Turn off the instrument, switch off the power supply and remove the power
cord.
b) Insert a small straight screwdriver into the slot at the power cord connector
and pry out the fuse holder gently.
c) Take out the fuse and replace it with a specified fuse. Then, install the fuse
holder.
7. AC Power Cord Connector
AC power input terminal. The power requirements of this oscilloscope are
100-240 V, 45-440 Hz. Use the power cord provided with the accessories to
connect the instrument to AC power. Then, you can press the Power key on the
front panel to start the instrument.