Chapter 5 To Trigger the Oscilloscope RIGOL
DS1000Z-E User Guide 5-31
Idle Time:
Press Idle to set the idle time before the edge counting in Nth edge trigger. The
range available is from 16 ns to 10 s.
Edge Number:
Press Edge to set the value of "N" in Nth edge trigger and the range available is from
1 to 65535.
Trigger Mode:
Press Sweep to open the trigger mode list and select auto, normal or single. For the
details, please refer to "Trigger Mode". The corresponding status light of the
current trigger mode turns on.
Trigger Setting:
Press Setting to set the trigger parameter (noise rejection) under this trigger type.
Trigger Level:
When the trigger source is an analog channel, you can use TRIGGER LEVEL to
modify the level. For the details, please refer to the introduction of "Trigger Level".