User Guide
Table Of Contents
- Guaranty and Declaration
- Safety Requirement
- DS2000 Series Overview
- Document Overview
- 1 Quick Start
- 2 To Set the Vertical System
- 3 To Set the Horizontal System
- 4 To Set the Sample System
- 5 To Trigger the Oscilloscope
- 6 To Make Measurements
- 7 Protocol Decoding
- 8 Reference Waveform
- 9 Pass/Fail Test
- 10 Waveform Record
- 11 Display Control
- 12 Store and Recall
- 13 System Function Setting
- 14 Remote Control
- 15 Troubleshooting
- 16 Specifications
- 17 Appendix
- Index
16 Specifications RIGOL
DS2000 User’s Guide
16-3
DS2102: 20 MHz
DS2072: 20 MHz
Low Frequency
Response
(AC Coupling, -3dB)
≤5 Hz (on BNC)
Calculated Rise
Time
1
DS2202: 1.8 ns
DS2102: 3.5 ns
DS2072: 5 ns
DC Gain Accuracy ±2% full scale
DC Offset Accuracy ±0.1 div ± 2 mV ± 1% offset value
Channel to Channel
Isolation
DC to maximum bandwidth: >40 dB
Trigger
Trigger Level Range
Internal ± 5 div from center of the screen
EXT ± 4 V
Trigger Mode Auto, Normal, Single
Holdoff Range 100 ns to 10 s
High Frequency
Rejection
1
75 kHz
Low Frequency
Rejection
1
75 kHz
Trigger Sensitivity 1 div (below 10 mV or noise rejection is enabled)
0.3 div (above 10 mV and noise rejection is disabled)
Edge Trigger
Edge Type Rising, Falling, Rising&Falling
Pulse Trigger
Pulse Condition Positive Pulse Width (greater than, lower than, within
specific interval)
Negative Pulse Width (greater than, lower than, within
specific interval)
Pulse Width Range 2 ns to 4 s
Runt Trigger
Pulse Width None, >, <, <>