Information

Chapter 16 Specifications RIGOL
DS4000E User’s Guide 16-1
Chapter 16 Specifications
All the specifications (except the parameters marked with "Typical") are guaranteed
when the instrument has been working for more than 30 minutes under the specified
operating temperature.
Sample
Sample Mode Real-time sample
Max. Real-time Sample
Rate
2.0 GSa/s for each channel
Max. Memory Depth 14 Mpts for each channel
Peak Detect 500 ps
Averaging After all the channels finish N times of sampling at the
same time, N can be 2, 4, 8, 16, 32, 64, 128, 256, 512,
1024, 2048, 4096, or 8192.
High Resolution 12 bits when 10 μs/div @ 2 GSa/s.
Input
Number of Channels 4 analog channels
Input Coupling DC, AC, or GND
Input Impedance (1 MΩ±1%) || (15 pF±3 pF) or 50 Ω±1.5%
Probe Attenuation
Coefficient
0.01X to 1000X, in 1-2-5 step
Maximum Input Voltage
(1 MΩ)
CAT I 300 Vrms, CAT II 100 Vrms, transient overvoltage
1000 Vpk
with RP2200 10:1 probe: CAT II 300 Vrms
with RP3300A 10:1 probe: CAT II 300 Vrms
with RP3500A 10:1 probe: CAT II 300 Vrms
with RP5600A 10:1 probe: CAT II 300 Vrms