RIGOL User’Guide DS6000 Series Digital Oscilloscope Oct. 2010 RIGOL Technologies, Inc.
RIGOL Guaranty and Declaration Copyright © 2010 RIGOL Technologies, Inc. All Rights Reserved. Trademark Information RIGOL is a registered trademark of RIGOL Technologies, Inc. Notices RIGOL products are protected by patent law in and outside of P.R.C. RIGOL reserves the right to modify or change parts of or all the specifications and pricing policies at company’s sole decision. Information in this publication replaces all previously corresponding materials.
RIGOL Safety Requirement General Safety Summary Please review the following safety precautions carefully before putting the instrument into operation so as to avoid any personal injuries or damages to the instrument and any product connected to it. To prevent potential hazards, please use the instrument as specified in this manual. Use Proper Power Cord. Only use power cords designed for the instrument and authorized by local country. Ground The Instrument.
RIGOL Avoid Circuit or Wire Exposure. Do not touch exposed junctions and components when the unit is powered. Do Not Operate With Suspected Failures. If you suspect damage occurs to the instrument, have it inspected by qualified service personnel before further operations. Any maintenance, adjustment or replacement of components must be performed by RIGOL authorized personnel. Keep Proper Ventilation. Inadequate ventilation may cause temperature increase which would damage the instrument.
RIGOL Safety Terms and Symbols Terms in this Manual. The following terms may appear in this manual: WARNING Warning statements indicate the conditions or practices that could result in injuries or loss of life. CAUTION Caution statements indicate the conditions or practices that could result in damage to this product or loss of data. Terms on the Product. The following terms may appear on the product: DANGER indicates a hazard may immediately happen. WARNING indicates potential hazard may happen.
RIGOL Measurement Category Measurement Category DS6000 series digital oscilloscopes can make measurements in Measurement Category I. WARNING This oscilloscope can only be used for measurements within its specified measurement categories. Measurement Category Definitions Measurement category I is for measurements performed on circuits not directly connected to MAINS. Examples are measurements on circuits not derived from MAINS, and specially protected (internal) MAINS derived circuits.
RIGOL Ventilation Requirement This oscilloscope uses fan to force cooling. Please make sure that the air intake and exhaust areas are free from obstructions and have free air. When using the oscilloscope in a bench-top setting, provide at least 10 cm clearance beside, above and behind the instrument for adequate ventilation. WARNING Inadequate ventilation may cause temperature increase which would damage the instrument.
RIGOL Working Environment Temperature Operating: 0℃ to +50℃ Non-operating: -20℃ to +70℃ Humidity Under +35℃: ≤90% relative humidity +35℃ to +40℃: ≤60% relative humidity WARNING To avoid short circuit inside the instrument or electric shock, please do not operate in humid environment. Altitude Operating: less than 3 km Non-operating: less than 15 km Installation (overvoltage) Category This product is powered by MAINS conforming to installation (overvoltage) category II.
RIGOL Pollution Degree Degree 2 Pollution Degree Definitions Pollution degree 1: No pollution or only dry, non-conductive pollution occurs. The pollution has no influence. For example: a clean room or air-conditioned office environment. Pollution degree 2: Normally only dry, non-conductive pollution occurs. Occasionally a temporary conductivity caused by condensation may occur. For example: general indoor environment.
RIGOL General Care and Cleaning General Care Do not leave the instrument exposed to direct sunlight for long period of time. Cleaning Clean the instrument regularly according to its use condition. To clean the oscilloscope: 1. Disconnect the instrument from power. 2. Clean the floating dust on the exterior of the instrument with a wet, drip-free and soft cloth (could use mild detergent or clear water). When cleaning an instrument with LCD, take care to avoid scratching the protective screen.
RIGOL Environmental Considerations The following symbol indicates that this product complies with the requirements of European Union according to Directives 2002/96/EC on waste electrical and electronic equipment (WEEE). Instrument Recycle Some substances contained in this product might be harmful to the environment or human health.
RIGOL DS6000 Series Overview Being a multifunctional and high performance digital oscilloscope, DS6000 delivers perfect combination of easy-to-use and powerful features to help users to fulfill their tasks more quickly. Main Features: 1 GHz or 600 MHz bandwidth. 2-channel or 4-channel model. 5 GSa/s maximum real-time sample rate, 100 GSa/s maximum equivalent sample rate, 180, 000 wfms/s (dots display) maximum waveform refresh rate. 140 Mpts maximum memory depth (standard).
RIGOL XII Provide shortcut keys for measurement, storage and print. Provide more than 2-hour lithium battery power supply to facilitate on-field test and use. Built-in 1 GBytes flash memory.
RIGOL Document Overview 1 Quick Start Describes the preparations before using the instrument and generally introduce the instrument. 2 To Set the Vertical System Introduces the functions of the vertical system of the oscilloscope. 3 To Set the Horizontal System Introduces the functions of the horizontal system of the oscilloscope. 4 To Set the Sample System Introduces the functions of the sample system of the oscilloscope.
RIGOL oscilloscope. 13 System Function Setting Introduces how to perform the remote interface, system-related and print operations. 14 Remote Control Introduces how to perform remote control of the oscilloscope. 15 Troubleshooting Introduces how to deal with common failures of the oscilloscope. 16 Specifications Lists the technical resolutions and general specifications of the oscilloscope. 17 Appendix Provides the options and accessories information.
RIGOL Contents Guaranty and Declaration .........................................................................I Safety Requirement ................................................................................ II General Safety Summary ........................................................................... II Safety Terms and Symbols ....................................................................... IV Measurement Category ......................................................................
RIGOL Knob ............................................................................................ 1-20 Navigation Knob ............................................................................ 1-21 Menu............................................................................................ 1-21 Record.......................................................................................... 1-22 MEAS ...........................................................................................
RIGOL Acquisition Mode ................................................................................... 4-2 Normal .......................................................................................... 4-2 Average ......................................................................................... 4-2 Peak Detect .................................................................................... 4-4 High Resolution ..............................................................................
RIGOL Advanced Operation....................................................................... 6-10 Auto Measurement ............................................................................... 6-13 Quick Measurement after Auto........................................................ 6-14 One-key Measurement of 22 Parameters .......................................... 6-15 Frequency Counter Measurement .................................................... 6-20 Measurement Setting .........................
RIGOL Waveform Playback .............................................................................. 10-4 11 Display Control............................................................................... 11-1 To Select the Display Type ..................................................................... 11-2 To Set the Persistence Time .................................................................. 11-3 To set the Waveform Brightness.............................................................
RIGOL Power Status ............................................................................... 13-14 External Trigger Impedance .......................................................... 13-14 Aux Output ................................................................................. 13-15 Reference Clock........................................................................... 13-16 Option Management .................................................................... 13-17 To Print the Waveform ...
RIGOL 1 Quick Start This chapter introduces the preparations when using the oscilloscope for the first time, the front panel, rear panel and user interface of the oscilloscope as well as the using method of the built-in help system.
RIGOL General Inspection 1. Inspect the shipping container for damage. If your shipping container appears to be damaged, keep the shipping container or cushioning material until you have inspected the contents of the shipment for completeness and have checked the oscilloscope electrically and mechanically. If your instrument has damaged during shipping, please contact your shipper and carrier for compensation. RIGOL will provide no free repair or replacement. 2. Inspect the instrument.
RIGOL Appearance and Dimensions DS6000 User’s Guide Figure 1-1 Front View Unit: mm Figure 1-2 Side View Unit: mm 1-3
RIGOL To Prepare the Oscilloscope for Use To Remove the Cover Before using the oscilloscope, remove the front panel cover by grasping the transverse grab on each side and pull them in the arrow directions as shown in the figure below.
RIGOL To Adjust the Supporting Legs Adjust the supporting legs properly to use them as stands to tilt the oscilloscope upwards, thus to place the oscilloscope stably and to operate and observe the display better.
RIGOL To Connect to AC Power Supply This oscilloscope accepts two kinds of AC power supply: 100-127V, 45-440Hz or 100-240V, 45-65Hz. Please use the power cord supplied with the accessories to connect the oscilloscope to the power source as shown in the figure below. After the power switch under the power plug is turned on, the oscilloscope is energized and the power key at the lower-left corner of the front panel is in breath state.
RIGOL To Use the Battery This oscilloscope also provides rechargeable battery on your option. Please install the battery as shown in the figure below. When leaving the factory, the battery holds certain electric quantity. When use the battery only, a battery icon will be displayed at the lower-right corner of the screen and the remaining electric quantity is indicated (no display when the AC power and battery are used at the same time).
RIGOL Power-on Inspection When the oscilloscope is energized, press the power key at the lower-left corner of the front panel to turn on the oscilloscope. During the start-up process, the oscilloscope performs a series of self-tests and you can hear the sound of relay switching. After the self-test, the start-up image is displayed. Press UTIL System SelfCalInfo to view the self-test results.
RIGOL To Connect the Probe RIGOL provides passive and active probes for the DS6000 series oscilloscopes. For detailed technical information of the probes, please refer to corresponding Probe User’s Guide. The following are the probes recommended for use with this oscilloscope. Model Description RP5600 600 MHz, passive probe, DS610X standard, auto detection RP3500 500 MHz, passive probe, DS606X standard, auto detection RP6150 1.5 GHz, passive probe, optional, auto detection RP7150 1.
RIGOL Function Inspection 1. Press SAVE Factory to restore the oscilloscope to its default configuration. 2. Use the probe to connect the input terminal of channel 1 (CH1) of the oscilloscope and the “Compensation Signal Output Terminal” of the probe. Compensation Signal Output Terminal Ground Terminal Figure 1-8 To Use the Compensation Signal 3. Connect the earth alligator clip of the probe to the “Ground Terminal” under the probe compensation signal output terminal. 4. Press the Auto key. 5.
RIGOL WARNING To avoid electric shock during the use of probe, please make sure that the insulated wire of the probe is in good condition and do not touch the metallic part of the probe when the probe is connected to high voltage source. Tip The signal output from the probe compensation connector can only be used for probe compensation adjustment and can not be used for calibration.
RIGOL Probe Compensation When the probes are used for the first time, you should compensate the probes to match the input channels of the oscilloscope. Non-compensated or poorly compensated probes may cause measurement inaccuracy and error. The probe compensation procedure is as follows: 1. Perform step 1, 2, 3 and 4 of “Function Inspection” in the previous section. 2. Check the displayed waveforms and compare them with the following figures.
RIGOL Front Panel Overview ① ② ③ ④ ⑤ ⑥⑦⑧ ⑨ ⑩ ⑪ ⑫ ⑬ ⑭ ⑮ ⑯ ⑰ ⑱ ⑲⑳ Figure 1-11 Front Panel Overview Table 1-1 Front Panel Description No. Description No.
RIGOL Rear Panel Overview 1. Trig Out/Calibration 3.Video Output 2. Reference Clock 5.USB DEVICE 4. LAN 7. AC Input/Switch 6.USB HOST 8. Lock Hole Figure 1-12 Rear Panel Overview 1. Trig Out/Calibration This connector can output various signals (press UTIL Aux Output to select the desired output type.): 1) TrigOut: After this type is selected, the oscilloscope output a signal that can reflect the current capture rate of the oscilloscope at each trigger.
RIGOL 2. Reference Clock Provide more precise sample clock signal for the oscilloscope and it can also synchronize two or more oscilloscope clocks. 3. Video Output Through this interface, the oscilloscope can be connected to external monitors such as projector to get clearer waveform display. Note that the display of the oscilloscope is still valid. 4. LAN Through this interface, the oscilloscope can be connected to the network for remote control.
RIGOL Front Panel Function Overview VERTICAL CH1, CH2, CH3, CH4: analog input channels. The four channels are marked by different colors which are also used to mark both the waveforms on the screen and the channel input connectors. Press any key to turn on the corresponding channel menu and press again to turn off the channel. AC: illuminated when AC channel coupling mode is enabled. 50: illuminated when the channel input impedance is 50Ω. BW: illuminated when the bandwidth limit is turned on.
RIGOL position quickly. VERTICAL SCALE: modify the vertical scale of the current channel. Turn clockwise to decrease the scale and turn counterclockwise to increase. During the modification, the amplitude of the waveform would enlarge or reduce and the scale information (e.g. ) at the lower side of the screen would change accordingly. Press down this knob to quickly switch the vertical scale adjustment modes between “Coarse” and “Fine”.
RIGOL HORIZONTAL MENU: press this key to turn on the horizontal control menu under which to turn on or off the delayed sweep function, switch between different time base modes, switch between “Coarse” and “Fine” of scale as well as modify the horizontal reference setting. HORIZONTAL SCALE: modify the horizontal time base. Turn clockwise to reduce the time base and turn counterclockwise to increase the time base.
RIGOL TRIGGER MODE: press this key to switch the trigger mode to Auto, Normal or Single and the corresponding state backlight of the current trigger mode would be illuminated. TRIGGER LEVEL: modify the trigger level. Turn clockwise to increase the level and turn counterclockwise to reduce the level. During modification, the trigger level line moves up and down and the value in the trigger level message box (e.g. ) at the lower-left corner of the screen changes accordingly.
RIGOL Single Press this key to set the trigger mode to “Single”. In single trigger mode, press the FORCE key to generate a trigger signal immediately. Auto Press this key to turn on the waveform auto setting function. The oscilloscope will automatically adjust the vertical scale, horizontal time base and trigger mode according to the input signal to realize optimum waveform display.
RIGOL Navigation Knob This knob provides quick Adjust/Locate function for numerical parameters with relatively large settable range. Turn clockwise (counter-clockwise) to increase (reduce) the value. Note that the inner knob is used for fine adjustment and the outer knob for coarse adjustment. For example, this knob can be used to quickly locate the waveform frame (“Current Frame” menu) to be played back in the waveform playback function.
RIGOL Record Play/Pause: press this key to play the recorded or paused waveform and press again to pause the play. The backlight is illuminated in yellow. Stop: press this key to stop the waveform in record or play back mode. The backlight is illuminated in orange. Record: press this key to start recording the waveform. The backlight is illuminated in red. MEAS Press this key to enter the measurement menu which supports measure setting, all measure, statistic and counter.
RIGOL User Interface DS6000 oscilloscope provides 10.2 inches, WVGA (800*480) 160,000 color TFT LCD, with ultra-wide screen, vivid picture, low power consumption and long service life. It supports standard display mode and persistence display mode. What is worth mentioning is that the 14-grid ultra-wide screen makes you view “longer” waveform.
RIGOL 1. Measurement Menu Provide 12 horizontal (HORIZONTAL) and 10 vertical (VERTICAL) measurement parameters. Press the softkey at the left of the screen to activate corresponding measurement item. 2. Channel Label/Waveform Different channels are marked by different colors and the color of the waveform complies with the color of the channel. 3. Status Available states include: RUN, STOP, T’D (triggered), WAIT and AUTO. 4.
RIGOL 9. Trigger Type Display the currently selected trigger type and trigger condition setting. Different labels are displayed when different trigger types are selected. For example: represents triggering on the rising edge in “Edge” trigger. 10. Trigger Source Display the trigger source currently selected (CH1-CH4, EXT, EXT/5 or AC Line). Different labels are displayed when different trigger sources are selected and the color of the trigger parameter area will change accordingly.
RIGOL The following labels will be provided according to the current channel setting: channel coupling, bandwidth limit and input impedance. 15. CH4 Vertical Scale Display the voltage value per grid of CH4 waveform vertically. Use VIRTICAL SCALE to modify this parameter. The following labels will be provided according to the current channel setting: channel coupling, bandwidth limit and input impedance. 16. Message Box Display prompt messages. 17.
RIGOL adjust/locate parameters. Denote that can be used to adjust parameters and then press to select the parameter. Denote that the current menu has several options. Denote that the current menu has a lower level menu. Press this key to return to the previous menu. Note that the following direction keys might appear in the grid at the lower-left corner of the menu bar: Denote that you can turn on the next page menu. Denote that you can turn on the previous page menu.
RIGOL To Use the Security Lock If needed, you can use the security lock (please buy it yourself) to lock the oscilloscope to a fixed location. The method is as follows, align the lock with the lock hole and plug it into the lock hole vertically, turn the key clockwise to lock the oscilloscope and then pull the key out.
RIGOL To Use the Desk Mount Instrument Arm Using an arm, the oscilloscope could be mounted on the work desk to save your operation space. The height and angle of the instrument could be adjusted freely to acquire supreme comfort and efficiency and to convenient your measurement and view. If needed, please buy and install corresponding option.
RIGOL To Use the Rack Mount Kit This oscilloscope can be mounted in a standard 19-inch rack cabinet that conforms to the Electric Industrial Association (EIA) standard. If needed, please buy and install the corresponding option. Kit Parts List No. Name Qty Part No.
RIGOL 2-1 2-2 2-3 Installation Tool PH2 Phillips Screwdriver (recommended).
RIGOL Installation Space The following requirements must be fulfilled by the machine cabinet in which the instrument is mounted. The machine cabinet must be a standard 19-inch one. At least 6U (266.7 mm) space should be provided by the machine cabinet. The depth inside the machine cabinet should not be less than 400 mm. The dimension of the instrument after being installed is shown below.
RIGOL Installation Procedures Only authorized operators can execute the installation operation. Improper installation might result in damage of the instrument or incorrect installation of the instrument on the rack. 1. Install the right and left plates: align the detents of the right and left plates with the openings on the support board and insert them into the support board respectively, then fix them with six M4 screws. 2.
RIGOL 3. Fix the instrument: fix the instrument on the support board with two back battens and twelve M4 screws. 4. Install the front panel: align the screw holes on the right and left plates with the screw holes on the front panel of the rack and let the front panel of the oscilloscope protrude from the opening. Then fix them with six M4 screws.
RIGOL 5. Load into the machine cabinet: mount the rack with the instrument fixed to it into a standard 19-inch machine cabinet with four M6 screws and four M6 square nuts. 6. Post-installation notice: the rack occupies a height of 6U. The holes pointed out by the arrows are installation holes. Note that they should be aligned with during installation.
RIGOL To Use the Built-in Help System The help system of this oscilloscope provides instructions for all the function keys (including menu keys) at the front panel. Press HELP to open the help interface and press again to close the interface. The help interface mainly consists of two parts. The left is “Help Options” and you can use “Button” or “Index” to select. The right is “Help Display Area”. Help Options Help Display Area Figure 1-15 Help Information Button: Default mode.
RIGOL 2 To Set the Vertical System The contents of this chapter: To Enable the Channel Channel Coupling Bandwidth Limit Probe Input Impedance Waveform Invert Vertical Scale Vertical Expansion Amplitude Unit Channel Label DS6000 User’s Guide 2-1
RIGOL To Enable the Channel DS6000 provides four analog input channels (CH1-CH4) and each channel can be controlled independently. CH1 Connect a signal to the channel connector of any channel for example CH1 and then press CH1 in the vertical control area (VERTICAL) at the front panel to enable CH1. At this point: Panel: The backlight of this key turns on. If the corresponding function in this menu has turned on, the characters “AC”, “50” or “BW” might also be illuminated.
RIGOL Channel Coupling Set the coupling mode to filter out the signals that is not needed. For example: the signal under test is a square waveform signal with DC offset. When the coupling mode is “AC”: the DC and AC components of the signal under test can both pass the channel. When the coupling mode is “AC”: the DC components of the signal under test are blocked. When the coupling mode is “GND”: the DC and AC components of the signal under test are both blocked.
RIGOL Bandwidth Limit Set the bandwidth to reduce display noise. For example: the signal under test is a pulse signal with high frequency oscillation. When the bandwidth limit is turned off, the high frequency components of the signal under test can pass the channel. Turn on the bandwidth limit and limit it to 20 MHz or 250 MHz, the high frequency components that exceed 20 MHz or 250 MHz are blocked. Press CH1 BW Limit and use to turn on/off the bandwidth limit (the default is off).
RIGOL Table 2-1 Probe Attenuation Coefficient Menu Attenuation coefficient 0.001X 1:1000 0.01X 1:100 0.1X 1:10 1X 1:1 10X 10:1 100X 100:1 1000X 1000:1 For “DiffProbe”, this menu blanks off. 3. Front When using “DiffProbe”, the differential probe always have different handle front accessories to be used to measure single-end signal and differential signal. Press this softkey to select “10Single” or “10Diff” according to the signal to be tested. 4.
RIGOL Input Impedance To reduce the circuit load caused by the interaction of the oscilloscope and the circuit under test, this oscilloscope provides two input impedance modes: 1 MΩ (default) and 50 Ω. 1 MΩ: at this point, the input impedance of the oscilloscope is very high and the current flows into the oscilloscope from the circuit under test can be ignored. 50 Ω: match the oscilloscope with devices whose output impedance is 50 Ω. Press CH1 Input to set the input impedance of the oscilloscope.
RIGOL Waveform Invert When waveform invert is turned on, the waveform display rotates 180 degree relative to the ground potential. When the waveform invert is turned off, the waveform display is normal. Press CH1 Invert to turn waveform invert on or off. Vertical Scale The vertical scale adjustment has two modes: “Coarse” and “Fine”. Press CH1 Volts/Div to select the desired mode. Turn the VERTICAL SCALE to adjust the vertical scale.
RIGOL Vertical Expansion When using VERTICAL SCALE to change the vertical scale of the analog channel, you can choose to expand or compress the vertical signal around the center of the screen or around the ground point of the signal. Press UTIL System VerticalExp to select “Center” or “Ground” and the default is “Ground”. Center: when the vertical scale is changed, the waveform will expand or compress around the center of the screen.
RIGOL Channel Label You can modify the labels used to mark the analog channels (CH1 to CH4) at the left side of the screen. The label is the number of the channel (e.g. ) by default and the length of the modified label can not exceed 4 characters. Press CH1 Label to enter the label modification interface. As shown in the figure below: Name Input Area For example: change “ Keyboard Upper-lower Case Switch ” to “ Input Method Switch ”. Press Keyboard to select the “Keyboard” area.
RIGOL 3 To Set the Horizontal System The contents of this chapter: Delayed Sweep Time Base Mode Horizontal Scale Horizontal Reference DS6000 User’s Guide 3-1
RIGOL Delayed Sweep Delayed sweep can be used to enlarge a length of waveform horizontally to view the details of the image. MENU Press the MENU key in the horizontal control area (HORIZONTAL) and then press the Delayed softkey to turn the delayed sweep on or off. Note that to turn on the delayed sweep, the current time base mode must be “Y-T” and the “Pass/Fail” is disabled. In delayed sweep mode, the screen is divided into two display areas as shown in the figure below.
RIGOL The waveform before enlargement Main time base Delayed Sweep 延迟扫描时基 The waveform after enlargement The waveform before enlargement: The waveform in the area that has not been covered by the subtransparent blue in the upper part of the screen is the waveform before enlargement. You can turn HORIZONTAL POSITION to move the area left and right or turn HORIZONTAL SCALE to enlarge or reduce this area.
RIGOL Time Base Mode Press the MENU key in the horizontal control area at the front panel and then press the Time Base softkey to select the time base mode of the oscilloscope and the default is Y-T mode. Y-T Mode This mode is the main time base mode and is applicable to CH1-CH4 input channels. In this mode, the Y axis represents voltage and the X axis represents time. Note that only when this mode is enabled can the “Delayed” be turned on.
RIGOL X-Y Mode In this mode, the oscilloscope automatically turns on all the four channels (CH1-CH4) and the screen is divided into two coordinate areas, where, X1 and Y1 track the voltages of CH1 and CH2 and X2 and Y2 track the voltages of CH3 and CH4. The phase deviation of two signals with the same frequency can be easily measured via Lissajous method. The figure below shows the measurement schematic diagram also called “Ellipse Method” of the phase deviation.
RIGOL X-Y function can be used to measure the phase deviation occurred when the test signal passes through a circuit network. Connect the oscilloscope to the circuit to monitor the input and output signals of the circuit. Application example: Measure the phase deviation of the input signals of two channels. Method 1: Use Lissajous method 1. 2. 3. 4. Connect a sine signal to CH1 and then connect a sine signal with the same frequency and amplitude but a 90° phase deviation to CH2.
RIGOL When X-Y mode is turned on, “Delayed” will be turned off automatically. The following functions are not available in X-Y mode: Auto measure, cursor measure, math operation, delayed sweep, vector display type, HORIZONTAL POSITION, trigger control, screen grid, memory depth, acquisition mode, Pass/Fail test and waveform recording. Method 2: Use the shortcut measurement function Please refer to “Phase A→B ” and “Phase A→B ” measurement functions of “Delay and Phase ” on page 6-17.
RIGOL Roll Mode In this mode, the waveform scrolls from the right to the left to update the display and the waveform horizontal position and trigger control are not available. Note that when Roll mode is turned on, “Delayed”, “Protocol Decoding”, “Pass/Fail” and “Meas.Range” are not available. Slow Sweep Another mode similar to the Roll mode.
RIGOL Horizontal Scale Being similar to “Vertical Scale”, the horizontal scale adjustment has two modes : “Coarse” and “Fine”. Press MENU ScaleAdjust in the horizontal control area (HORIZONTAL) at the front panel to select the desired mode. Turn HORIZONTAL SCALE to adjust the horizontal scale. Turn clockwise to reduce the horizontal scale and turn counterclockwise to increase the scale. The scale information (e.g. ) at the upper left corner of the screen will change accordingly during the adjustment.
RIGOL Horizontal Reference Horizontal reference is the reference position according to which the screen waveform expands and compresses horizontally when adjusting HORIZONTAL SCALE. In Y-T mode, press MENU HorRef in the horizontal control area (HORIZONTAL) to select the desired reference mode and the default is “Center”. 1. The Center of the Screen When changing the horizontal time base, the waveform expands or compresses horizontally around the center of the screen. 2.
RIGOL DS6000 User’s Guide 3-11
RIGOL 4 To Set the Sample System The contents of this chapter: Acquisition Mode Sample Mode Sample Rate Memory Depth Antialiasing DS6000 User’s Guide 4-1
RIGOL Acquisition Mode The acquisition mode is used to control how to generate waveform points from the sample points. Press ACQ Acquisition in the function menu at the front panel and use to select the desired acquisition mode (the default is normal), then press the knob to select this mode. You can press the Acquisition softkey continuously to switch the acquisition mode. ACQ Normal In this mode, the oscilloscope samples the signal at equal time interval to rebuild the waveform.
RIGOL The available range of the number of averages is from 2 to 8192 and the default is 2. When “Average” mode is selected, press the Averages menu and use to set the desired number of averages as the power function of 2.
RIGOL Peak Detect In this mode, the oscilloscope acquires the maximum and minimum values of the sample interval signal to get the envelope or the narrow pulse of the signal that might be lost. Use this mode to avoid signal confusion at the expense of exaggerating any noise. In this mode, the oscilloscope can display all the pulses whose pulse widths are at least as wide as the sample period.
RIGOL Sample Mode Press ACQ Sampling and press this softkey continuously to switch the sample mode (the default is real-time sample). Note that all the “Acquisition Mode” support real-time sample and equivalent sample. Real-time Sample In this mode, the oscilloscope samples and displays waveform within a trigger event. The maximum real-time sample rate of DS6000 is 5 GSa/s and the current sample rate is displayed under the Sa Rate menu.
RIGOL Sample Rate Sample rate refers to the process that the oscilloscope converts the analog signals to discrete data at a certain time interval and stores the data in order. Sample rate is the reciprocal of the time interval. Sample Rate 1 /Δ t The sample rate of this oscilloscope is up to 5 GSa/s.
RIGOL 3. Waveform Leakage: when the sample rate is too low, the waveform rebuilt by the sample data does not reflect all the actual signal information.
RIGOL Memory Depth Memory depth refers to the number of waveform points that the oscilloscope can store in a single trigger sample and it reflects the storage ability of the acquisition memory. DS6000 provides a memory depth up to 140 M points.
RIGOL Antialiasing At slower sweep speeds, the sample rate is reduced and a proprietary display algorithm is used to minimize the likelihood of aliasing. Press ACQ Anti_aliasing to turn on or off the antialiasing function. By default, Antialiasing is disabled. The displayed waveforms will be more susceptible to aliasing when this function is disabled.
RIGOL 5 To Trigger the Oscilloscope When working, the oscilloscope continuously acquires waveform data even if a valid trigger does not occur. However, only when valid triggers occur, can the display become stable. The trigger circuit guarantees that each sweep or sample starts from the position of the trigger event in the input signal, namely each sweep and sample are synchronous and the waveforms acquired overlap to display stable waveform.
RIGOL Trigger Source Press MENU Source in the trigger control area (TRIGGER) at the front panel to select the desired trigger source. Signals input from the analog channels CH1-CH4 and the [EXT TRIG] connector as well as the AC Line can all be used as the trigger source. MENU Analog channel input: Signals input from analog channels CH1-CH4 can all be used as the trigger source. No matter whether the input from the channel selected is displayed, the channel can work normally.
RIGOL Trigger Mode Trigger mode would influence the mode in which the oscilloscope searches for trigger event. The following is the schematic diagram of the acquisition memory from which you can see that the position of the trigger event is determined by the reference time point and the delay setting. Trigger Event Pre-trigger Buffer Post-trigger Buffer Acquisition memory Pre-trigger/ Delayed trigger: Acquire data before/after the trigger event.
RIGOL Auto: No matter whether the trigger condition is met, there is always waveform display. A horizontal line is displayed when no signal is input. When this mode is selected, the oscilloscope fills the pre-trigger buffer first and then search for a trigger while at the same time continues filling data. When the oscilloscope is searching for trigger, the data firstly filled into the pre-trigger buffer will overflow and be pushed out of the FIFO queue.
RIGOL Trigger Coupling Trigger coupling decides which kind of components will be transmitted to the trigger circuit. Note to distinguish it with “Channel Coupling”. DC: allow the DC and AC components to pass the trigger path. AC: block all the DC components and attenuate signals lower than 8 Hz. LF reject: block the DC components and reject the low frequency components lower than 8 kHz. HF reject: reject the high frequency components higher than 65 kHz.
RIGOL Trigger Holdoff Trigger holdoff could trigger complicated waveforms (e.g. pulse sequence) stably. Holdoff time is the time the oscilloscope waits to re-enable the trigger circuit. During the holdoff time, the oscilloscope will not trigger before the end of the holdoff time. Trigger position Holdoff Time Press MENU Set Up Holdoff in the trigger control area (TRIGGER) at the front panel and use to change the holdoff time (the default is 100 ns) until the waveform triggers stably.
RIGOL Trigger Type DS6000 has abundant trigger functions including multiple serial bus triggers.
RIGOL Edge Trigger Trigger on the trigger threshold of the specified edge of the input signal. Trigger Type: Press Type to select “Edge”. At this point, the trigger setting information as shown in the figure below is displayed at the upper right corner of the screen. Source Selection: Press Source to select CH1-CH4, EXT, EXT/5 or AC Line as the Trigger Source. The current trigger source is displayed at the upper right corner of the screen.
RIGOL Pulse Trigger Trigger on positive or negative pulse with specified width. Trigger Type: Press Type to select “Pulse”. At this point, the trigger setting information as shown in the figure below is displayed at the upper right corner of the screen. Source Selection: Press Source to select CH1-CH4, EXT or EXT/5 as the Trigger Source. The current trigger source is displayed at the upper right corner of the screen. Pulse Condition: Press When to select the desired pulse condition.
RIGOL A B Trigger level Positive pulse width When the Pulse Condition is set to , , or , press Set Up and use to input the desired value. The range available is from 4 ns to 4 s. When the Pulse Condition is set to or , press Upper Limit and Lower Limit respectively and use to input the desired values. The range of the upper limit available is from 12 ns to 4 s. The range of the lower limit available is from 4 ns to 3.99 s.
RIGOL Slope Trigger Trigger on the positive or negative slope with specified time. Trigger Type: Press Type to select “Slope”. At this point, the trigger setting information as shown in the figure below is displayed at the upper right corner of the screen. Source Selection: Press Source to select CH1-CH4, EXT or EXT/5 as the Trigger Source. The current trigger source is displayed at the upper right corner of the screen. Slope Condition: Press When to select the desired slope condition.
RIGOL A B The upper limit of trigger level (Up Lvl) The lower limit of trigger level (Low Lvl) Positive Slope time When the Slope Condition is set to , , or , press Time and use to input the desired value. The range available is from 10 ns to 1 s. When the Slope Condition is set to or , press Upper Limit and Lower Limit respectively and use to input the desired values. The range of time upper limit available is from 10 ns to 1 s. The range of the time lower limit available is from 10 ns to 1 s.
RIGOL unchanged. : adjust the upper and lower limits of the trigger level at the same time. During the adjustment, the “UP Lvl” and the “Low Lvl” change accordingly but the “Slew Rate” remains unchanged. Trigger Mode: Press Sweep to select the Trigger Mode (page 5-3) under this trigger type as auto, normal or single. The corresponding status light of the current trigger mode turns on.
RIGOL Video Trigger Trigger on the standard video signal fields and lines of NTSC (National Television Standards Committee), PAL (Phase Alternating Line) or SECAM (Systeme Electronique Pour Couleur Avec Memoire). Trigger Type: Press Type to select “Video”. At this point, the trigger setting information as shown in the figure below is displayed at the upper right corner of the screen. Source Selection: Press Source to select CH1-CH4, EXT or EXT/5 as the Trigger Source.
RIGOL SECAM: the frame frequency is 25 frames per second. The sweep line is 625 with interlacing sweep. Trigger Mode: Press Sweep to select the Trigger Mode (page 5-3) under this trigger type as auto, normal or single. The corresponding status light of the current trigger mode turns on. Trigger Setting: Press Set Up to set the trigger parameters (trigger coupling and trigger holdoff) under this trigger type. Trigger Level: Use TRIGGER LEVEL to modify the level.
RIGOL Pattern Trigger Identify the trigger condition through searching for the specified pattern. Pattern is the logic “AND” combination of each channel and the value of each channel can be H (High), L (Low) or X (Don’t Care). One channel in the pattern can be specified with rising edge or falling edge and if the pattern settings of all the other channels are “True” (H or L), the oscilloscope will trigger on the specified edge.
RIGOL define another edge item, thus the former edge item defined will be ignored. Trigger Mode: Press Sweep to select the Trigger Mode (page 5-3) under this trigger type as auto, normal or single. The corresponding status light of the current trigger mode turns on. Trigger Setting: Press Set Up to set the trigger parameters (trigger coupling, trigger sensitivity and trigger holdoff) under this trigger type. Trigger Level: Use TRIGGER LEVEL to modify the level.
RIGOL RS232 Trigger Trigger according to the start frame, error frame, check error and data. Trigger Type: Press Type to select “RS232”. At this point, the trigger setting information as shown in the figure below is displayed at the upper right corner of the screen. Source Selection: Press Source to select CH1-CH4 as the Trigger Source. The current trigger source is displayed at the upper right corner of the screen. Trigger Condition: Press When to select the desired trigger condition.
RIGOL transmission. The rates available include: 2400 bps, 4800 bps, 9600 bps (default), 19200 bps, 38400 bps, 57600 bps, 115200 bps and user-defined. After “User” is selected, press the Set Up softkey and use to set more specific value with the adjustment step as 1 bps. Trigger Mode: Press Sweep to select the Trigger Mode (page 5-3) under this trigger type as auto, normal or single. The corresponding status light of the current trigger mode turns on.
RIGOL I2C Trigger Trigger on the start condition, restart, stop and missing acknowledgement or on the read/write frame with specific device address and data value. When using I2C trigger, you need to specify the SCL and SDA data sources. Trigger Type: Press Type to select “I2C”. At this point, the trigger setting information as shown in the figure below is displayed at the upper right corner of the screen. Source Selection: Press SCL and SDA to specify the data sources of SCL and SDA respectively.
RIGOL --press the Qualifier softkey and select “=”, “>” or “<”. A&D: trigger when the “Address” and “Data” conditions are met at the same time. After this trigger condition is selected: --press the AddrBits softkey to select “7 bit” or “10 bit”; -- press the Address softkey to set the address value according to the setting in AddrBits and the ranges are from 0 to 127 and from 0 to 1023 respectively; -- press the Data softkey and input the data value.
RIGOL SPI Trigger Trigger on the data pattern with specified edge. When using SPI trigger, you need to specify the SCL, SDA and CS data sources. Trigger Type: Press Type to select “SPI”. At this point, the trigger setting information as shown in the figure below is displayed at the upper right corner of the screen. Source Selection: Press SCL, SDA and CS to specify the data sources of SCL, SDA and CS respectively. CH1-CH4 can be selected as the Trigger Source.
RIGOL Press ClockEdge to select the desired clock edge. : trigger on the rising edge of the data pattern when the voltage level meets the preset Trigger Level. : trigger on the falling edge of the data pattern when the voltage level meets the preset Trigger Level. Trigger Mode: Press Sweep to select the Trigger Mode (page 5-3) under this trigger type as auto, normal or single. The corresponding status light of the current trigger mode turns on.
RIGOL CAN Trigger Trigger on the specified frame type of the data frame. When using CAN trigger, you need to specify the signal source, signal rate and trigger signal type of the CAN signal. Trigger Type: Press Type to select “CAN”. At this point, the trigger setting information as shown in the figure below is displayed at the upper right corner of the screen. Source Selection: Press Source to select CH1-CH4, EXT or EXT/5 as the Trigger Source.
RIGOL Sample Point: Sample point is the point within the bit time. The oscilloscope samples the bit level at this point to determine whether the bit level is explicit or implicit. Sample rate represents the time percentage of the start and end of the bit time. Press SamplePoint and use to modify the parameter with the step as 1% and the range is from 5% to 95%. 1Bit 60% 70% 80% Trigger Mode: Press Sweep to select the Trigger Mode (page 5-3) under this trigger type as auto, normal or single.
RIGOL USB Trigger Trigger on the SOP, EOP, RC, Suspended and Exit Suspended of the data packet on the differential USB data cable (D+ and D-). This trigger supports USB Low Speed and Full Speed. Trigger Type: Press Type to select “USB”. At this point, the trigger setting information as shown in the figure below is displayed at the upper right corner of the screen. Source Selection: Press D+ and D- to specify data sources for D+ and D- data cables respectively.
RIGOL Use TRIGGER LEVEL to modify the level. For details, please refer to the description of “Trigger Level” on page 5-8.
RIGOL Trigger Ouput Connector The trigger output connector at the rear panel can output trigger signals decided by the current setting. Trigger Output Connector Press UTIL AuxOutput to select “TrigOut”. When the oscilloscope is triggered, it will output a trigger signal decided by the current setting through the [Trig Out/Calibration] interface.
RIGOL 6 To Make Measurements After data is sampled and displayed, the DS6000 can make math operation, cursor measurement and auto measurement.
RIGOL Math Operation DS6000 can realize various math operations of waveforms between channels including: addition (A+B), subtraction (A-B), multiplication (AxB), division (A÷ B), FFT, logic operation and advanced operation. The results of math operation also allows further measurement (for details, please refer to “Cursor Measurement”). MATH Press MATH Operate in the vertical control area (VERTICAL) at the front panel to select the desired operation function.
RIGOL Subtraction Subtract the waveform voltage values of signal source B from A point by point and display the results. Press MATH Operate to select “A-B”: Press Source A and Source B to select the desired channels. The channels available are CH1, CH2, CH3 and CH4. Press Invert to turn the invert display of the operation results on or off. VERTICAL POSITION, VERTICAL SCALE, HORIZONTAL POSITION and HORIZONTAL SCALE can also be used to adjust the display of the operation results.
RIGOL Division Divide the waveform voltage values of signal source A by B point by point and display the results. It can be used to analyze the multiple relationships of waveforms in two channels. Note that when the voltage value of channel B is 0, the result of the division is treated as 0. Press MATH Operate to select “A ÷ B”: Press Source A and Source B to select the desired channels. The channels available are CH1, CH2, CH3 and CH4.
RIGOL FFT FFT is used to quickly perform Fourier transform on specified signals and transform time domain signals to frequency domain signals. Use FFT to facilitate the following works: Measure harmonic components and distortion in the system Measure the characteristics of the noise in the DC power Analyze vibration Press MATH Operate to select “FFT” and set the parameters of FFT operation.
RIGOL 2. Select Window Function Spectrum leakage can be considerably decreased when a window function is used. DS6000 provides four kinds of FFT window functions with different characteristics and are applicable to measure different waveforms. You need to select the window function according to different waveforms and characteristics. Press Window to select the desired window function and the default is “Rectangle”. Table 6-1 Window Function 3.
RIGOL Full Screen: the source channel and the FFT operation results are displayed in the same window to view the frequency spectrum more clearly and to perform more precise measurement. Note that in FFT mode and when MATH is the active channel, you can also press HORIZONTAL SCALE to switch to “Split” or “Full Screen”. 4. Set the Vertical Scale In FFT measurement, the unit of the horizontal axis changes from time to frequency.
RIGOL Logic Operation Perform logic operation on the waveform voltage values of the specified source point by point and display the results. In operation, when the voltage value of the source channel is greater than 0, it is regarded as logic “1”, otherwise logic “0”.
RIGOL Press MATH Operate to select “Logic”: Press Expression to select the desired operation expression and the default is AND. Press Source A and Source B to select the desired channels. The channels available are CH1, CH2, CH3 and CH4. Press Invert to turn the invert display of the operation results on or off. VERTICAL POSITION, VERTICAL SCALE, HORIZONTAL POSITION and HORIZONTAL SCALE can also be used to adjust the display of the operation results.
RIGOL Advanced Operation DS6000 provides advanced operation function that allows user-defined operation functions. Press MATH Operate Advance Expression On, and the editing window as shown in the figure below pops up on the screen. Turn to select any item ( if they are currently available for selection) in “Channel”, “Function”, “Variable”, “Operator” and “Figure”, then press the knob and the item selected will de displayed in the entry box on the right of the “Expression”.
RIGOL 1. Expression Here, it refers to the formulas formed by channel, function, variable, operator and figure. The length of the expression is limited to 64 bytes. 2. Channel You can select one or more channels from CH1, CH2, CH3 and CH4. Function Please refer to the following table to get the functions of each function. Note that the left brackets “(” here are only used to facilitate your entry and they are not a part of the function name. 3. Table 6-3 Function 4.
RIGOL variable. After setting the “Step” or pressing this softkey, turn the knob to modify this parameter. The range available is from -9.9999 to 9.9999. Exponent: press this softkey to set the numeric values of the exponents with 10 as the mantissa in the variables. The range is from -9 to 9. For example, the following setting means set Variable1 to 6.1074× 108. Variable: Variable1 Mantissa: 6.1074 Exponent: 8 5. Operator Please refer to the following table to get the functions of each operator.
RIGOL Auto Measurement DS6000 provide auto measurements of 22 waveform parameters and the statistic and analytical functions of the measurement results. What’s more, you can also use the frequency counter to realize more precise frequency measurement.
RIGOL Quick Measurement after Auto When the oscilloscope is correctly connected and has detected input signal, press Auto to display the waveform automatically and open the following function menu: Single-period: measure the “Period” and “Frequency “of the current signal within a single period and display the measurement results at the bottom of the screen.
RIGOL One-key Measurement of 22 Parameters Press the MENU key at the left of the screen to turn on the measurement menu of the 22 parameters and then press the corresponding menu softkey to quickly realize “One-key” measurement. The measurement result will be displayed at the bottom of the screen. The icons of time and voltage parameters in the measurement items and the measurement results on the screen are always marked in the same color according to the channel (MEAS Source) currently used.
RIGOL Time Parameters RiseTime FallTime 90% 50% 10% +Width 1. 2. 3. 4. 5. 6. 7. 8. 6-16 -Width Period: defined as the time between the two crossing points of the middle threshold values of two continuous edges with the same polarity. Frequency: defined as the reciprocal of period. Rise Time: the time that the signal amplitude rises from 10% to 90%. Fall Time: the time that the signal amplitude falls from 90% to 10%.
RIGOL Delay and Phase Period Source A Delay Source B 1. Delay A→B : the time deviation between the rising edges of source A and 2. source B. Negative delay means that the rising edge of source A appears later than that of source B. Delay A→B : the time deviation between the falling edges of source A and 3. 4. source B. Negative delay means that the falling edge of source A appears later than that of source B.
RIGOL Voltage Parameters Overshoot Vmax Vtop Vamp Vpp Vbase Vmin Preshoot 1. 2. 3. 4. 5. 6. 7. Vmax: the voltage value from the highest point of the waveform to the GND. Vmin: the voltage value from the lowest point of the waveform to the GND. Vpp: the voltage value from the highest point to the lowest point of the waveform. Vtop: the voltage value from the flat top of the waveform to the GND. Vbase: the voltage value from the flat base of the waveform to the GND.
RIGOL 9. Overshoot: the ratio of the difference of the maximum value and top value of the waveform to the amplitude value. 10. Preshoot: the ratio of the difference of the minimum value and base value of the waveform to the amplitude value.
RIGOL Frequency Counter Measurement The hardware frequency counter supplied with this oscilloscope can make more precise measurement of the signal frequency. Press MEAS Counter to select anyone of CH1-CH4 as the source. The measurement result is displayed at the upper right corner of the screen and you can identify the current measurement source according to the color of the icon. The following figure is the result of the frequency measurement of the signal from CH1.
RIGOL Measurement Setting 1. Source Selection Press MEAS Source to select the desired channel for measurement (CH1-CH4). The color of the parameter icon under the MENU at the left of the screen will change with the source selected. 2. Measurement Range Press MEAS Meas.Range Meas.Range to select “Screen Region” or “Cursor Region” for measurement. When “Cursor Region” is selected, two cursor lines appear on the screen.
RIGOL Press MEAS Meas Setting Type “Phase” and then press Source A and Source B to set the two channel sources (CH1-CH4) of the phase measurement respectively. 5. Threshold Measurement Setting Specify the vertical level (in percentage) being measured in the analog channel. Measurement of all the time, delay and phase parameters will be influenced by this setting. Press MEAS Meas Setting Type “Threshold” and then: Press Max and use to set the maximum value of the measurement.
RIGOL Clear the Measurement If you have currently turned on one or more items in the 22 measurement parameters, you can “Delete” or “Recover” the 5 parameters ahead or “Delete” or “Recover” all the measurement items that have been turned on. Note that the 5 parameters ahead is determined according to the order in which you turned them on and they will not change as you delete one or more measurement items. Press MEAS Clear Item n to “Delete” or “Recover” the specified measurement item.
RIGOL All Measurement All measurement could measure all the time and voltage parameters (totally 18) of the current measurement source and display the results at the bottom of the screen. Press MEAS Display All to turn the function on or off. When all measurement is turned on, “One-key” measurement is also valid. “Clear” will not clear all the measurement results.
RIGOL Statistic Function Make statistic and display the current, average, minimum (or standard deviation) and maximum (or count) values of at most 5 measurement items that are turned on last. Press MEAS Statistic to turn the statistic function on or off. Press MEAS StatisSel to select “Extremum” or “Difference” measurement. When “Extremum” is selected, minimum and maximum values are displayed. When “Difference” is selected, standard deviation and count values are displayed.
RIGOL Measurement History To view the history measurement data, press MEAS MeasHistory MeasHistory “On”. The history data can be displayed in two modes: Graph: display the results of the last 10 measurements of at most 5 measurement items that are turned on last in graph mode. The measurement points are connected using linear interpolation. Table: display the results of the last 10 measurements of at most 5 measurement items that are turned on last in table mode.
RIGOL Cursor Measurement Cursors are the horizontal and vertical marks that can be used to measure the X axis values (usually Time) and Y axis values (usually Voltage) on a selected waveform. Please connect the signal to the oscilloscope and obtain stable display before using cursor measurement. All the “Auto Measure” parameters can be realized through cursor measurement. Press the front panel function menu CURS Mode, use to select the desired cursor mode (the default is OFF) and then press the knob.
RIGOL Manual Mode In this mode, a pair of cursors will appear. You can adjust the cursors to measurement the X (or Y), X increment (or Y increment), and reciprocal of the X increment on the selected source (CH1-CH4 or MATH).
RIGOL 3. Select X (Y) Axis Unit When the cursor type is “Time”, press Time Unit to select “s”, “Hz”, “°” or “%”. S: when this unit is selected, in the measurement results, the CurA, CurB and △X are in “s” and 1/△X is in “Hz”. Hz: when this unit is selected, in the measurement results, CurA, CurB and △X are in “Hz” and 1/△X is in “s”. °: when this unit is selected, in the measurement results, CurA, CurB and △X are in “°”.
RIGOL 5. Measurement Example Use manual cursor measurement to measure the period (△X) of a square waveform and the result is 1 ms equal to the result from auto measurement.
RIGOL Track Mode In this mode, one or two pairs of cursors will appear. You can adjust the two pairs of cursors to measure the X and Y values on two different sources respectively. The points being measured on cursor A and B are marked by an orange rectangle and rhombus respectively. When the cursors are moved horizontally, the marks will position on the waveform automatically.
RIGOL operation results (MATH) as the measurement source of cursor B (only channels that are turned on are available). You can also select “None”, namely do not use cursor B. 2. Adjust Cursor Position Adjust cursor A: press CursorA and use to adjust the position of cursor A. During the adjustment, the measurement result will change accordingly. The adjustable range is limited within the screen. Adjust cursor B: press CursorB and use to adjust the position of cursor B.
RIGOL Cursor Track (Before Horizontal Expansion): Cursor Track (After Horizontal Expansion): DS6000 User’s Guide 6-33
RIGOL Auto Mode In this mode, one or more cursors will appear. You can use auto cursor measurement to measure anyone of the 22 waveform parameters. Before using this mode, you need to at least turn on one auto measurement parameter and the number of cursors will change with the measurement parameters. Press CURS Mode “Auto” and the cursor corresponding to the measurement parameter which is turned on firstly will be displayed on the screen.
RIGOL 7 Protocol Decoding Protocol analysis can be used by users to discover errors, debug hardware and accelerate development easily, so as to guarantees quick and high-quality accomplishment of projects. Protocol decoding is the basis of protocol analysis. Only protocol analyses with correct protocol decoding are acceptable and only correct protocol decoding can provide more error information.
RIGOL Parallel Decoding Parallel bus consists of clock line and data line. As shown in the figure below, CLK is the clock line, while Bit0 and Bit1 are the 0 bit and 1 bit on the data line respectively. CLK Bit0 Bit1 The oscilloscope will sample the channel data on the rising edge, falling edge or the rising&falling edges of the clock and judge whether each data point is logic “1” or logic “0” according to the preset threshold level.
RIGOL Next, press Channel to specify a channel source for the bit currently selected in ChanBits. 3. Analog Channel Threshold Setting To judge logic “1” and logic “0” of the buses, you need to set a threshold for each analog channel (CH1, CH2, CH3 and CH4). When the signal amplitude is greater than the preset value, it is considered as “1”, otherwise “0”. Threshold Level 1 0 0 Press Threshold to turn the threshold setting menu on. 4. Channel Select the channel that needs to set a threshold.
RIGOL RS232 Decoding (Option) RS232 serial bus consists of the transmitting data line (TX) and the receiving data line (RX). Rx Tx DeviceA DeviceB Tx Rx The industry standard of RS232 uses “Negative Logic”, namely high level is logic “0” and low level is logic “1”. 0 0 1 0 1 1 1 0 1 1 By default, RS232 uses LSB (Least Significant Bit) transmission sequence, namely the lowest bit of the data is transmitted first.
RIGOL Stop Bit Check Bit Start Phase Data Bits Start Bit: represent when the data starts outputting. Setting the Polarity is equivalent to having specified the “Start Bit”. Data Bits: represent the number of data bits actually contained in each frame of data. Even-OddCheck: be used to check the correctness of the data transmission. Odd-Check: the sum of the number of “1” in the data bit and check bit is an odd.
RIGOL 5. Data Packet Setting As mentioned before, in RS232, you need to set the start bit, the data bits, the check bit (optional) and the stop bit of each frame of data. “Start Bit” has been specified by the “Polarity Setting”. The setting methods of other parameters are introduced below: Press Data Bits to set the data width of each frame. It can be set to 5, 6, 7, 8 or 9 and the default is 8. Press Stop Bit to set the stop bit after each frame of data. It can be set to 1 bit, 1.5 bits or 2 bits.
RIGOL Check Error: When check errors are detected during the decoding, the data will be filled with red batches.
RIGOL SPI Decoding (Option) SPI serial bus consists of chip select line (SS), clock line (SCLK), MISO and MOSI. SS SS Host SS MISO MISO MOSI MOSI SCLK SCLK SCLK Device MISO MOSI SS: select positive or negative polarity. When positive polarity is selected, low level is valid. When negative polarity is selected, high level is valid. When SS selection is valid, the SPI bus samples data from MISO and MOSI at the jumping point of SCLK.
RIGOL edge of SCLK. Press Threshold to set the threshold of the clock channel. 3. MISO Setting Press MISO to turn the MISO data line setting interface. Press Channel to select any channel (CH1-CH4) as the MISO channel. When “OFF” is selected, this data line is not set. Press Polarity to set the polarity of the MISO data line to positive polarity or negative polarity. Press Threshold to set the threshold of the MISO data line. 4.
RIGOL 7-10 DS6000 User’s Guide
RIGOL I2C Decoding (Option) I2C serial bus consists of the clock line (SCLK) and the data line (SDA). Vcc A2 SCLK SDA Device A1 A0 Host SCLK SDA SCLK: sample the SDA on the clock rising edge or falling edge. SDA: denote the data channel. Press MATH BUS1 Decode to select “I2C” and turn the I2C decoding function menu on. 1. SCLK Setting Press SCLK to select any channel (CH1-CH4) as the clock channel. Press SCLKThreshold to set the threshold of the clock channel. 2.
RIGOL Press Offset and use to adjust the vertical display position of the bus. Press BusStatus to turn the bus display on or off. 4. 7-12 Error Expressions during Decoding In I2C bus, the front part of each frame of data contains the address information and blue patches are used to represent address ID. In the ID, “W” is used to represent writing address and “R” is used to represent reading address. When the ACK (ACKnowledge Character) is not met, the data block is filled with red patches.
RIGOL 8 Reference Waveform In actual testing process, the waveform being tested can be compared with the reference waveform module to judge the reasons for failures.
RIGOL To Enable REF Function Press the REF function key in the vertical control area (VERTICAL) at the front panel to enable REF function. Note that when the time base is in X-Y mode, REF function can not be enabled. REF DS6000 provides 10 reference waveform channels. Press Channel and use to select to turn the desired reference channel on or off and a channel icon of the channel that is turned on will be display at the left side of the screen grid. For example: .
RIGOL To Select REF Source Press Current and use to select any reference channel (Ref1-Ref10) that has been turned on and then press Source to specify a reference source (CH1-CH4, MATH) for this channel. To Save to Internal Memory Press Save to save the waveform (screen region) in the specified source to internal memory as reference waveform and display it on the screen. Note that this operation only saves the reference waveform to the volatile memory and it is not saved after power-off.
RIGOL To Export to Internal or External Memory Users can also save the reference waveform to the Flash memory inside the instrument or to the external USB flash device. The file format of the reference waveform is “*.ref”. At most 10 reference files (LocalREF1.ref to LocalREF10.ref) can be saved inside the instrument. Press Export to enter file saving interface. Please refer to the relative descriptions in “Store and Recall” to save the reference waveform to internal or external memory.
RIGOL 9 Pass/Fail Test Monitor the change of the signal by judging whether the input signal is within the creation mask. The monitor information can be displayed on the screen as well as be prompted by the system sound or the pulse signal output from the [Trig Out/Calibration] connector at the rear panel.
RIGOL To Enable Pass/Fail Test Function Press UTIL Pass/Fail to turn on the Pass/Fail test menu. Note that when the time base is in X-Y mode, the Pass/Fail test function can not be enabled. UTIL To start testing, press Operate. Select “ ” to start testing and select “■” to stop testing. The figure below is the test interface: You can select the parameters such as Source, Create Mask, Save and Load the Test Mask. For details, please refer to the following introduction.
RIGOL To Select Source Press Source to select the channel (CH1-CH4) to be tested. During the test, the oscilloscope will judge whether each frame of waveform in the source complies with the current test mask and those waveforms pass through the mask area is considered as failed. To Create Mask Users can define their own test masks. Press Enable Test to turn the mask display area on. Press X Mask and Y Mask respectively and turn , the mask lines appear on the screen.
RIGOL are detected. The test results on the screen will update continuously and the [Trig Out/Calibration] connector at the rear panel will output a pulse each time a failed waveform is detected. Press Output to select “Fail” or “Fail+ ”. Fail: when failed waveforms are detected, there are display and output but the beeper does not sound. Fail+ : when failed waveforms are detected, there are display and output and the beeper sounds (not related to the on/off state of the sound).
RIGOL 10 Waveform Recording Waveform recording can not only record the waveforms in the input channels (CH1-CH4) but also record the passed or failed waveforms in the Pass/Fail test. Better waveform analysis effect can be reached with the playback functions.
RIGOL Waveform Recording Waveforms from all the channels currently turned on will be recorded during waveform recording. WaveFinder (Playback) UTIL Play/Pause (Playback) Stop Recording/Playback Record (Record) Press UTIL Record Mode and use waveform recording operation menu on. 1. 2. End Frame Press End Frame and use to set the desired number of recording frames. Recording Operation Waveform recoding can be realized via the shortcut buttons in the menu or at the front panel.
RIGOL 4. Maximum End Frame The menu shows the maximum end frame that can be recorded currently. As the capacity of the waveform memory is fixed, the more the number of points of each frame of waveform has, the less the number of waveform frames can be recorded. Thus, the maximum end frame of waveform recording is decided by the “Memory Depth”. Please refer to the instruction in “Memory Depth” to select the desired memory depth.
RIGOL Waveform Playback Waveform playback can play back the waveforms currently recorded or the stored waveforms recalled from the memory. Press UTIL Record Mode and use to select “Play Back”. At this point, the information as shown in the figure below is displayed at the upper left corner of the screen, indicating the specific frame currently displayed on the screen. During the playback, this value would change continuously. Please refer to the following descriptions to set the playback parameters.
RIGOL 5. End Frame Press this softkey and use to set the end frame of the playback. The default is the number of frames of the waveform to be played back. 6. Playback Operation Waveform playback can be realized through the shortcut buttons in the menu or at the front panel. Menu Press Operate and select “ start playing back. Front Panel ” to Press Operate and select “ ” to pause the playback.
RIGOL 11 Display Control You can set the type, the persistence time and the brightness of waveform display as well as the grid type, grid brightness of the screen display and the menu display time.
RIGOL To Select the Display Type DISP Press DISP Type to set the waveform display mode to “Vectors” or “Dots”. 11-2 Vectors: the sample points are displayed in connecting line mode. Normally, this mode can provide the most vivid waveform to view the steep edge of the waveform (e.g. square waveform). Dots: display the sample points directly. You can directly view each sample point and use the cursor to measure the X and Y values of the sample point.
RIGOL To Set the Persistence Time Press DISP Persis. Time to set the persistence time of the oscilloscope to Vmin, specific values (From 50 ms to 20 s) or Infinite. In the following part, a frequency sweep signal of the sine waveform is used to demonstrate the waveform effects in different persistence times. 1. Vmin Enable to view the waveform changing in high refresh rate. 2. Specific Values Enable to observe glitch that changes relatively slowly or glitch with low occurrence probability.
RIGOL 3. 11-4 Infinite In this mode, the oscilloscope displays the waveform newly acquired without clearing the waveforms acquired formerly. The waveforms acquired formerly will be displayed in relatively low-brightness color and the waveform newly acquired will be displayed in normal brightness and color. Infinite persistence can be used to measure noise and jitter and to capture incidental events.
RIGOL To set the Waveform Brightness Press DISP WaveIntensity or turn in normal operation mode to adjust the waveform brightness of the analog channel. The default is 50% and the range available is from 0% to 100%. To Set the Screen Grid Press DISP Grid to set the screen grid type. : turn the background grid and coordinate on. : turn the background grid off. : turn the background grid and coordinate off.
RIGOL 12 Store and Recall Users can save the current settings, waveforms, and screen image in internal memory or external USB mass storage device (e.g. USB flash device) in various formats and recall the stored settings or waveforms when needed.
RIGOL Storage System Press SAVE to enter the store and recall setting interface. SAVE Insert the USB flash device This oscilloscope provides two USB Host interfaces at the front and rear panel to connect USB flash storage device for external storage. The USB flash storage device connected is marked with “Disk E” (front panel) and “Disk D” (rear panel).
RIGOL Storage Type Press SAVE Storage to select the desired storage type. The default is “Traces”. The store and recall descriptions of each type are as follows: 1. Traces Save the data in internal or external memory in “*.trc” format. Internal storage: only save the data of the selected channel, and at recall, the data will be displayed on the screen directly. At most 10 trace files (from LocalTrace0.trc to LocalTrace9.trc) can be stored.
RIGOL 5. 12-4 CSV Save the waveform data of all the channels turned on in external memory in a single “*.csv” file. You can specify the file name and the saving directory and save the corresponding parameter file (*.txt) under the same directory using the same file name. The recall of CVS and parameter files is not supported.
RIGOL Internal Store and Recall Internal store and recall support “Traces”, “Waveforms” and “Setups” in Storage. In the following part, “Traces” are taken as an example for illustration. 1. Save the specified type of file in internal memory. 1) Connect the signal to the oscilloscope and obtain stable display. 2) Press SAVE Storage to select “Traces” and press Save to turn on the interface as shown in Figure a. use to select “ Local Disk” and press to turn on the local disk (Figure b).
RIGOL Figure b Note that in internal storage, New menu is not available. 2. Load the specified type of file in the internal memory. 1) Press SAVE Storage to select “Traces” and then press Load to turn on the interface as shown in Figure c. Use to select “Local Disk” and then press to turn on the local disk (Figure d). Figure c 2) As shown in Figure d, use to select the desired file to load and press Load to load the file selected.
RIGOL Figure d DS6000 User’s Guide 12-7
RIGOL External Store and Recall Before using external store and recall, make sure that a USB flash device is connected correctly. External storage supports all the types of files in Storage but in recall, “Picture” and “CSV” are not supported. 1. Save the specified type of file in the external USB flash device. 1) Connect the signal to the oscilloscope and obtain stable display. 2) Press SAVE Storage to select “CSV” and press Save to turn on the interface as shown in Figure e.
RIGOL Figure f 4) After the storage position is selected, press New to turn on the interface as shown in Figure g. Refer to the descriptions in “To Create a New File or Folder” to create a new file name. Figure g 5) Press OK to execute the saving operation. 2. Load the specified type of file in the external USB flash storage device. 1) Press SAVE Storage to select “Traces” and then press Load to turn on the interface as shown in Figure h.
RIGOL Figure h 2) As shown in Figure i, use to select the desired file to load and then press Load to load the selected file.
RIGOL Disk Management Press SAVE Disk.Manage to turn on the disk management interface as shown in the figure below and use to select the desired disk. The disk currently selected is displayed in green and press to turn on the selecte diskd. Execute the following operations through the disk management menu: To Select File Type To Create New, Delete or Rename Files or Folders To Copy and Paste Files.
RIGOL To Create a New File or Folder This operation is only valid in external storage. Before using external store and recall, make sure that a USB flash device is connected correctly. First, press SAVE Disk Manage and use to select and turn on the external disk (“Disk D” or “Disk E”). Second, select the desired directory under which to create a new file or folder. The default is the root directory of the USB flash device. Next, press New to turn on the interface as shown in the figure below.
RIGOL English Input Method For example: create a file or folder using the name “Filename”. 1. Press Keyboard. 1) Use to select English input method “En” and uppercase input state “aA”. 2) Use to input the letter “F”. If the input is wrong, press Delete to delete the character input. 3) Use to select lowercase input state “Aa”. 4) Use to input the remaining letters “ilename”. Name Input Area 2. 3.
RIGOL Chinese Input Method For example: create a file or folder using the name “文件名”. 1. Press Keyboard. 1) Use to select Chinese input method “中” and lowercase input state “ Aa”. Note that the “Chinese” menu is added in the menu items at the right of the screen. 2) Use to input the pinyin “wen”. If the input is wrong, press Delete to delete the pinyin input. After “wen” is input, a series of Chinese characters appear in the “Chinese Chraracter Selecting Area”.
RIGOL To Delete a File or Folder Folder operation is valid only in external storage. Before using the external disk, make sure that a USB flash device is connected correctly. 1. Delete a file in internal memory. 1) Press SAVE Disk Manage and use to select and turn on the local disk (“local Disk”). 2) Press Type to select the desired file type to delete. 3) Use to select the desired file to delete. 4) Press Delete to delete the file selected. 2. Delete a file or folder in external memory.
RIGOL To Copy and Paste a File You can copy a file in the local disk to the external disk or copy a file in one external disk to another external disk. Before using the external disk, make sure that a USB flash device is connected correctly. Press SAVE Disk Manage and use to select and turn on the local disk (“local Disk”). Use to select the desired file to copy and then press Copy to copy the selected file.
RIGOL Factory Press SAVE Factory to return the oscilloscope to its factory state. Refer to the table below.
RIGOL Bandwidth OFF Input Impedance 1 MHz Channel Invert OFF Vertical Scale Coarse Channel Unit V Acquisition Setting (ACQ) Acquisition Mode Normal Sampling Mode Real Time Memory Depth Auto Trigger Setting (TRIGGER) Trigger Type Edge Source CH1 Slope Rising Edge Trigger Mode Auto Trigger Coupling DC Trigger Holdoff 100 ns Display Setting (DISP) Display Type Vectors Persistence Time Min Waveform Intensity 50% Screen Grid Full Brightness 50% Menu Display Infinite Curso
RIGOL Track Cursor A CH1 Cursor B CH2 CurA -4*1 us CurB 4*1 us Storage Setting (SAVE) Storage Type Traces Utility Function Setting (UTIL) I/O Setting Network Configuration Mode DHCP, Auto IP USB Device Auto Detect Sound Sound OFF Pass/Fail Test Enable Test OFF Source CH1 Operate OFF X Mask 0.48 Y Mask 0.
RIGOL Power On Recall Last Vertical Expansion Ground Screen Saver Default Screen Saver Time OFF External Trigger Impedance 1 MΩ Aux TrigOut Reference Clock None Math Operation Setting (MATH->Operation) Operate OFF A+B Source A CH1 Source B CH1 Invert OFF Scale 2V A-B Source A CH1 Source B CH1 Invert OFF Scale 5 uV A*B Source A CH1 Source B CH1 Invert OFF Scale 5U A/B Source A CH1 Source B CH1 Invert OFF Vertical Scale 1U FFT Source CH1 Window Function Rect
RIGOL Vertical Scale 1Vrms Vertical Scale 1.
RIGOL Packet End NULL SPI SS Channel None SS Polarity Positive SCLK Channel CH2 SCLK Slope Rising Edge MISO Channel CH1 MISO Channel Positive MOSI Channel CH1 MOSI Polarity Positive MOSI Data Bits 8 MOSI Endian MSB I2C SCLK CH1 SDA CH2 Reference Waveform Setting (REF) Channel Setting REF1 Current Channel REF1 Source CH1 12-22 DS6000 User’s Guide
RIGOL 13 System Function Setting The contents of this chapter: Remote Interface Configuration System-related To Print the Waveform DS6000 User’s Guide 13-1
RIGOL Remote Interface Configuration DS6000 can communicate with PC via LAN, USB and GPIB (together with the USB to GPIB module of RIGOL) buses. Please refer to the introduction below to configure the corresponding interface before using the remote interfaces. LAN Setting Press UTIL IO Setting LAN Set to turn on the LAN setting interface. You can view the network connection status and configure the network parameters.
RIGOL IP Configuration Type (DHCP) The configuration type of the IP address can be DHCP, auto IP or static IP. In different IP configuration type, the configuration mode of the network parameters such as the IP address is different. Press Config Mode and use to select “DHCP”. Then press to select this type. When DHPC type is valid, the DHCP server will assign the network parameters such as the IP address for the oscilloscope. IP Configuration Type (Auto IP) Press Config Mode and use to select “Auto IP”.
RIGOL nnn is from 0 to 223 (except 127) and the ranges of the other three nnn are from 0 to 255. You are recommended to ask your network administrator for an IP address available. Press IP Address and use to input the desired IP address. This setting will be saved in the non-volatile memory and if DHCP and Auto IP are set to “OFF”, the oscilloscope will load the preset IP address automatically at the next power-on. 2. Set the Subnet Mask. The format of the subnet mask is nnn.nnn.nnn.
RIGOL range of the first nnn is from 0 to 223 (except 127) and the ranges of the other three nnn are from 0 to 255. You are recommended to ask your network administrator for an address available. press DNS and use to input the desired address. Generally, users do not need to set the DNS, therefore this parameter setting can be ignored. Tips When the three IP configuration types are all turned on, the priority of the parameter configuration from high to low is “DHCP”, “Auto IP” and “Static IP”.
RIGOL USB Device This oscilloscope can communicate with a PC or PictBridge printer via the USB Device interface at the rear panel. You need to set the oscilloscope to make it match with different device types. Press UTIL IO Setting USB and use to select the desired device type. Auto Detect: the oscilloscope can detect the device currently connected to the USB Device interface automatically. Computer: under this type, the oscilloscope can communicate with a PC.
RIGOL System-related Sound When the sound is enabled, you can hear the sound of the beeper when you press a function key or a menu softkey or when the prompt message pops up. Press UTIL Sound to select sound is turned on, a trumpet icon (on) or (off). The default is off. When the will be displayed at the lower right corner of the screen. Language This oscilloscope supports multiple language menus, Chinese/English help and prompt message. Press UTIL Language and use to select the desired language.
RIGOL Power-off Recall You can set the system configuration to be recalled when the oscilloscope is powered on again after power-off. Press UTIL System Power On to select “Last” (default) or “Default”. Last: return to the setting of the system at last power-off. Default: return to the factory setting of the system. System Time The system time is displayed at the lower right corner of the screen in “mm:ss (minute: second)” format.
RIGOL Year: 2000 to 2099 Month: 01-12 Date: 01-31 (28, 29 or 30) Hour: 00 to 23 Minute: 00 to 59 Second: 00 to 59 Press Apply to put the current setting into effect. The time at the lower right corner of the screen will be updated.
RIGOL Self-calibration Information Press UTIL System SelfCalInfo to view the result of the last self-calibration of the oscilloscope. The self-calibration result usually contains the contents as shown in the figure below. Note: If the oscilloscope is installed with battery, the self-calibration information will also display the electric quantity, voltage and current of the battery.
RIGOL Screen When the oscilloscope enters the idle state and holds for a certain period of time, the screen saver program will be enabled. Press UTIL System Screen to turn on the screen saver setting menu. “Default” means using the icon RIGOL as the screen saver icon. Press Time to select the time of the screen saver. When “OFF” is selected, the screen saver program is disabled. The time of the screen saver can be set to: 1 min, 2 min, 5 min, 15 min, 30 min, 45 min, 60 min, 2 hour or 5 hour.
RIGOL Error Information If this menu is shown in gray, it means that the circuit board of the oscilloscope is working normally. Otherwise, an exclamation mark will be displayed in the status bar at the lower right corner of the screen, indicating that there is error information currently and at this point, press this key (press UTIL System ErrorInfo) to view. The error information that might appear including: the temperature of the mainboard is too high, the fan stops rotating etc.
RIGOL Self-calibration Self-calibration program can quickly help the oscilloscope to reach the best working state and get the most precise measurement values. You can perform self-calibration at any time especially when the change range of the environment temperature is up to or more than 5 ℃. Make sure that the oscilloscope has been warmed up or operated for more than 30 minutes.
RIGOL Power Status Users can set the power status of the oscilloscope after power-on. Connect the oscilloscope to the AC source using the power cord and turn on the power switch at the rear panel, the oscilloscope is powered on. Press UTIL PowerStatus to select “Default” or “Open”. Default: after the oscilloscope is powered on, you need to press the power key at the front panel to start up the oscilloscope. Open: after the oscilloscope is powered on, it starts up directly.
RIGOL Aux Output Users can set the signal type output from the [Trig Out/Calibration] connector at the rear panel. Press UTIL AuxOutput to select the desired output type. 1. TrigOut After this type is selected, the oscilloscope output a signal that can reflect the current capture rate of the oscilloscope at each trigger. 2. Quick Edge After this type is selected, the oscilloscope can output a quick edge signal which can be used in the self-calibration of the oscilloscope. 3.
RIGOL Reference Clock This oscilloscope can output the internal 10 MHz sample clock from the [10MHz In/Out] connector at the rear panel and accept an external 10 MHz clock to synchronize multiple oscilloscopes. 10MHz In/Out Press UTIL RefClock to select the desired clock type. ClockOutput: configure the [10MHz In/Out] connector as an output and output the internal 10 MHz clock.
RIGOL Option Management This oscilloscope provides multiple options to fulfill your measurement requirements. Please contact your RIGOL sales representative or RIGOL technical support to order the corresponding options. You can view the options currently installed on the oscilloscope or activate the newly bought option serial number through the menu. Press UTIL Options Installed to view the options currently installed in the oscilloscope.
RIGOL To Print the Waveform To Connect the Print Device DS6000 oscilloscope supports normal and PictBridge printers and you can set the print parameters to print the current screen image. The connection modes of different print devices are different. Use the USB Host interface to connect normal printer and use the USB Device interface to connect the PictBridge printer.
RIGOL Normal Print After “Normal” is selected, please refer to the following steps to set the print parameters to execute print. 1. Print Range Press this softkey to select “Screen” (default) or “Wave”. Screen: print the whole screen image. Wave: only print the waveform and the basic information. 2. Palette Press this softkey to select “Grayscale” (default) or “Color”. 3. Invert Press this softkey to turn the invert print on or off. 4. Preview Press this softkey to turn on the preview interface.
RIGOL PictBridge Print PictBridge is a new print standard. If your oscilloscope and the printer both comply with the PictBridge standard, you can connect the oscilloscope to the printer using a USB cable to print the screen image directly. Devices complies with the PictBridge standard always have the icon as shown in the figure on the right. After selecting the “PictBridge” print mode, please refer to the following steps to set the print parameters and execute print. 1.
RIGOL function. 7. Quality Press this softkey to select the print quality. You can select: Default, Normal, Draft or Fine. Note that the selection is unavailable when the printer does not support this function. 8. Date Press this softkey to turn the date print on or off. Note that this attribute is the built-in attribute of the PictBridge printer. This menu is invalid when the printer connected does not support this attribute. 9. Invert Press this softkey to turn the invert print on or off. 10.
RIGOL 14 Remote Control The oscilloscope can be controlled remotely mainly using the following two methods: User-defined programming Users can program and control the oscilloscope by using the SCPI (Standard Commands for Programmable Instruments) commands. For more information about the commands and programming, refer to the Programming Guide.
RIGOL Remote Control via USB 1. Connect the device Connect the oscilloscope (USB DEVICE) with your PC using a USB cable. 2. Install the USB driver This oscilloscope is a USBTMC device. Assuming that your PC has already been installed with Ultra Sigma, after you connect the oscilloscope to the PC and turn both on for the first time, the PC will display a New Hardware Wizard as shown in the figure below. Please install the “USB Test and Measurement Device” driver following the directions in the wizard.
RIGOL 5 6 7 DS6000 User’s Guide 14-3
RIGOL 3. Search device resource Start up the Ultra Sigma and the software will automatically search for the oscilloscope resources currently connected to the PC. You can also click to search the resources. During the search, the status bar of the software is as shown in the figure below: 4.
RIGOL DS6000 User’s Guide 14-5
RIGOL Remote Control via LAN 1. Connect the device Connect the oscilloscope to your LAN using a network cable. 2. Configure network parameters Configure the network parameters of the oscilloscope according to the description in “System Function Setting Remote Interface Configuration LAN Setting”. 3. Search device resource Start up the Ultra Sigma and click , then the software will search for the oscilloscope resources currently connected to the LAN.
RIGOL 5. Communication test Right click the resource name “DS6104 (TCPIP::172.16.3.88::INSTR)” to select “Instrument Common RIGOL SCPI ControlPanel V1” to turn on the remote control command control panel, namely you can use this panel to send commands and read data. As shown in the figure below: 6. Load LXI webpage As this oscilloscope conforms to LXI-C standards, you can load LXI webpage through Ultra Sigma (right-clicking the resources name and selecting LXI-Web).
RIGOL 14-8 DS6000 User’s Guide
RIGOL Remote Control via GPIB 1. Connect the device Use the USB to GPIB module to extend a GPIB interface for the oscilloscope. Then connect the oscilloscope to your PC using a GPIB cable. 2. Install the driver of GPIB card Install the driver of the GPIB card which has been connected to the PC correctly. 3. Set the GPIB address Set the GPIB address of the oscilloscope according to the description in “System Function Setting Remote Interface Configuration To Set the GPIB Address”. 4.
RIGOL problem. 5. View device resource Click “OK” to return back to the main interface of Ultra Sigma. The resources found will appear under the “RIGOL Online Resource” directory. 6. Communication Test Right-click the resource name “DS6104 (GPIB0::18::INSTR)” to select “Instrument Common RIGOL SCPI ControlPanel V1” to turn on the remote control command control panel, namely you can use this panel to send commands and read data.
RIGOL DS6000 User’s Guide 14-11
RIGOL 15 Troubleshooting The commonly encountered failures and their solutions are listed below. When you encounter those problems, please solve them following the corresponding steps. If the problem remains still, please contact RIGOL and provide your device information (acquisition method: Utility System Sys Info). 1. The screen still dark (no display) after power on: (1) Check if the power is correctly connected or if the battery is correctly installed. (2) Check if the power switch is really on.
RIGOL (2) Check the trigger type: general signals should use “Edge” trigger and video signal should use “Video” trigger. Only when the proper trigger type is used, can the waveform be displayed stably. (3) Try to change the Coupling to “HF Reject”or “LF Reject” to filter out the high-frequency or low-frequency noise that disturbs the trigger. (4) Change the trigger holdoff setting. 5.
RIGOL 16 Specifications All the specifications are guaranteed except the parameters marked with “Typical” and the oscilloscope needs to operate for more than 30 minutes under the specified operation temperature. Sample Sample Mode Real-time Sample, Equivalent Sample Real Time Sample Rate 5 GSa/s (single-channel) 2.
RIGOL Probe Attenuation Coefficient 0.001X, 0.01X, 0.
RIGOL 230 mV/div to 5 V/div: ± 40V (1MΩ) 2 Bandwidth Limit 20 MHz or 250 MHz Low Frequency Response (AC Coupling -3dB) ≤5 Hz (on BNC) Calculated Rise Time2 DS606X: 600 ps DS610X: 400 ps DC Gain Accuracy ± 2% full scale DC Offset Accuracy 200 mV/div to 5 V/div: 0.1 div ± 2 mV± 0.5 offset value 2 mV/div to 195 mV/div: 0.1 div ± 2 mV± 1.
RIGOL Negative Slope (greater than, lower than, within specific interval) Time Setting 10 ns to 1 s Video Trigger Signal Standard Line Frequency Range Support standard NTSC, PAL and SECAM broadcasting standards, the range of the number of lines is from 1 to 525 (NTSC) and 1 to 625 (PAL/SECAM) Pattern Trigger Pattern Setting H, L, X, Rising Edge, Falling Edge RS232/UART Trigger Trigger Condition Start, Error, Check Error, Data Baud Rate 2400bps, 4800bps, 9600bps, 19200bps, 38400bps, 57600bps, 11520
RIGOL Trigger condition SOP, EOP, RC, Suspended, Exit Suspended Measure Cursor Manual Mode Auto Measurement Voltage Deviation between Cursors (△ V) Time Deviation between Cursors (△ T) Reciprocal of △ T (Hz) (1/△ T) Track Mode Voltage and Time Values of the Waveform Point Auto Mode Allow to display cursors during auto measurement Measurements of Maximum, Minimum, Peak-Peak Value, Top Value, Bottom Value, Amplitude, Average, Mean Square Root, Overshoot, Pre-shoot, Frequency, Period, Rise Time, Fall
RIGOL FFT Vertical Scale Linear RMS, dBV RMS Logic Operation AND, OR, NOT, XOR Math Function Intg, Diff, Log, Exp, Sqrt, Sine, Cosine, Tangent Number of Buses for Decoding 2 Decoding Type Parallel (standard), RS232/UART (option), I2C&SPI (option) Display Display Type 10.
RIGOL 100-240 V/45-65 Hz Power Maximum 150W Fuse 3 A, T Degree, 250 V Environment Temperature Range Operating: 0 ℃ to +50 ℃ Non-operating: -20 ℃ to +70 ℃ Cooling Method Humidity Range fan cooling Under +35 ℃: ≤90% Relative Humidity +35 ℃ to +50 ℃: ≤60% Relative Humidity Altitude Operating: under 3,000 meters Non-operating: under 15,000 meters Physical Characteristics Size3 Width× Height× Depth = 399.0 mm× 255.3 mm× 123.8 mm 4 Weight Package Excluded 5.345 ± 0.2 kg Package Included 10.
RIGOL 17 Appendix Appendix A: Options and Accessories Model Standard Accessories Optional Accessories Decoding Options Description Order Number DS6104 (1 GHz, 4-channel) DS6104 DS6102 (1 GHz, dual-channel) DS6102 DS6064 (600 MHz, 4-channel) DS6064 DS6062 (600 MHz, dual-channel) DS6062 Power Cord conforming to the standard of the country - Front Panel Cover FPC-DS-6 USB Cable CB-USB-150 2 or 4 Passive Probes (600 MHz) RP5600 (for DS610X) 2 or 4 Passive Probes (500 MHz) RP3500 (for DS
RIGOL Appendix B: Warranty RIGOL warrants that its products mainframe and accessories will be free from defects in materials and workmanship within the warranty period. If a product is proven to be defective within the respective period, RIGOL guarantees the free replacement or repair of products which are approved defective. For detailed warranty description, please refer to RIGOL official website or the warranty card.
RIGOL Appendix C: Any Comment or Question? If you have any comment or question on our document, please send email to: service@rigol.
RIGOL Index Acquisition Mode ........................ 4-2 Addition ..................................... 6-2 AND .......................................... 6-8 Auto IP .................................... 13-3 Auto Mode ............................... 6-34 Aux Output ............................ 13-15 Average ..................................... 4-2 BandWidth Limit ......................... 2-4 Blackman ................................... 6-6 CAN Trigger..............................
RIGOL Trigger Coupling ......................... 5-5 Trigger Holdoff ........................... 5-6 Trigger Level .............................. 5-8 Trigger Mode .............................. 5-3 Trigger Position ........................ 3-10 Trigger Sensitivity ....................... 5-6 Trigger Source ............................ 5-2 USB......................................... 13-6 USB Trigger.............................. 5-26 Variable ................................... 6-11 Vectors ...............