2-7
Waveform Invert
When waveform invert is turned on, the waveform display rotates 180 degree
relative to the ground potential. When the waveform invert is turned off, the
waveform display is normal. Press CH1 Invert to turn waveform invert on or off.
Vertical Scale
The vertical scale adjustment has two modes: “Coarse” and “Fine”.
Press CH1 Volts/Div to select the desired mode. Turn the VERTICAL
SCALE to adjust the vertical scale. Turn clockwise to reduce the scale and turn
counterclockwise to increase the scale.
The scale information (e.g. ) in the channel label at the bottom of the
screen will change accordingly during the adjustment. The adjustable range of the
vertical scale is from 2 mV/div to 5 V/div.
Coarse adjustment (take counterclockwise as an example): set the vertical scale
in 1-2-5 step namely 2 mV/div, 5 mV/div, 10 mV/div5 V/div.
Fine adjustment: further adjust the vertical scale within a relatively smaller
range to improve vertical resolution. If the amplitude of the input waveform is a
little bit greater than the full scale and the amplitude would be a little bit lower if
the next scale is used, fine adjustment can be used to improve the amplitude of
waveform display to view the details of the signal.
Note that you can not only use the Volts/Div menu to select “Coarse/Fine” but also
press VERTICAL SCALE to quickly switch.