User manual

RIGOL Chapter 2 Command System
2-70 MSO7000/DS7000 Programming Guide
:DVM:MODE
Syntax
:DVM:MODE <mode>
:DVM:MODE?
Description
Sets or queries the mode of the digital voltmeter.
Parameter
Name
Type
Range
Default
<mode>
Discrete
{ACRMs|DC|DCRMs}
ACRMs
Description
ACRMs: displays the root-mean-square value of the acquired data, with the DC
component removed.
DC: displays the average value of the acquired data.
DCRMs: displays the root-mean-square value of the acquired data.
Return
Format
The query returns ACRM, DC, or DCRM.
Example
:DVM:MODE DC /*Sets the mode of the digital voltmeter to DC.*/
:DVM:MODE? /*The query returns DC.*/