Data Sheet

RIGOL
4
Displayed Average Noise Level (DANL)
0 dB RF Attenuation, RBW=100 Hz, VBW=1Hz,
RMS Average Detector, Trace Average ≥ 50, Input
Impedance=50 Ω, Tracking Generator Off.
DANL
(Preamplier Off)
100 kHz to 10 MHz:
<-85 dBm-3 x (f/1 MHz) dB, typical
-125 dBm
10 MHz to 2.5 GHz:
<-127 dBm+3 x (f/1GHz) dB, typical
-130 dBm
2.5 GHz to 3 GHz:
<-115 dBm
100 kHz to 10 MHz:
<-75 dBm-3 x (f/1 MHz)
dB, typical -115 dBm
10 MHz to 2.5 GHz:
<-117 dBm+3 x (f/1 GHz)
dB, typical -120 dBm
2.5 GHz to 3 GHz:
<-105 dBm
100 kHz to 10 MHz: <-75
dBm-3 x (f/1 MHz) dB,
typical -115 dBm
10 MHz to 2 GHz:
<-117 dBm+3 x (f/1 GHz)
dB, typical -120 dBm
DANL
(Preamplier On)
100 kHz to 1 MHz:
<-103 dBm
1 MHz to 10 MHz:
<-103 dBm-3 x (f/1 MHz) dB, typical
-143 dBm
10 MHz to 2.5 GHz:
<-145 dBm+3 x (f/1 GHz) dB, typical
-148 dBm
2.5 GHz to 3 GHz:
<-133 dBm
100 kHz to 1 MHz:
<-93 dBm
1 MHz to 10 MHz:
<-93 dBm-3 x (f/1 MHz)
dB, typical -133 dBm
10 MHz to 2.5 GHz:
<-135 dBm+3 x (f/1 GHz)
dB, typical -138 dBm
2.5 GHz to 3 GHz:
<-123 dBm
Level Display Range
Log Scale 1 dB to 200 dB
Linear Scale 0 to Reference Level
Number of Display Points Normal: 601; Full Screen: 751
Number of Traces 3 + Math Trace
Trace Detectors Normal, Positive-peak, Negative-peak, Sample, RMS, Voltage Average, Quasi-Peak
Trace Functions Clear Write, Max Hold, Min Hold, Average, Freeze, Blank
Scale Units dBm, dBmV, dBμV, V, W
Frequency Response
10 dB RF Attenuation, Relative to 50 MHz, 20
to 30
Frequency Response
Preamplier Off , 100 kHz to 3 GHz:
<0.7 dB
Preamplier On , 1 MHz to 3 GHz:
<1.0 dB
Preamplier Off , 100
kHz to 3 GHz: <1.0 dB
Preamplier On , 1 MHz
to 3 GHz: <1.4 dB
100 kHz to 2 GHz: <1.0 dB
Input Attenuation Switching Uncertainty
Setting Range 0 to 50 dB, in 1 dB step
Switching Uncertainty (fc=50 MHz,
relative to 10 dB, 20 ºC to 30 ºC)
< (0.3 + 0.01 x attenuator setting)
dB
<0.8 dB
Absolute Amplitude Uncertainty
Uncertainty
(fc=50 MHz, peak detector,
preamplier off, 10 dB RF
attenuation, input signal=-10 dBm,
20 ºC to 30 ºC)
±0.4 dB
RBW Switching Uncertainty
10 Hz to 1 MHz, relative
to 1 kHz RBW
100 Hz to 1 MHz, relative to 1 kHz RBW
Uncertainty <0.1 dB
Reference Level
Range -100 dBm to +30 dBm, in 1 dB step
Resolution Log Scale: 0.01 dB; Linear Scale: 5 digits
Level Measurement Uncertainty
95% condence level, S/N>20 dB, RBW=VBW=1 kHz, preamplier off, 10 dB RF attenuation, -50 dBm<reference level<0,10
MHz<fc<2 GHz(DSA1020), 10 MHz<fc<3 GHz(DSA1030A/DSA1030A-TG/DSA1030/DSA1030-TG), 20 ºC to 30 ºC
Level Measurement Uncertainty <1.0 dB, nominal <1.5 dB, nominal
0 dB RF Attenuation, RBW=10
Hz, VBW=1Hz, RMS Average
Detector, Trace Average ≥ 50, Input
Impedance=50 Ω, Tracking Generator
Off.