User manual

RIGOL Chapter 2 Front Panel Operation
User’s Guide for DSA700 Series
2-70
Mkr->Stop
Set the stop frequency of the analyzer to the frequency of the current marker.
If Normal marker is selected, the stop frequency will be set to the frequency
of the current marker.
If Delta, Delta Pair or Span Pair marker is selected, the stop frequency will
be set to the frequency of the Delta Marker.
The function is invalid in Zero span mode.
Mkr->Ref
Set the reference level of the analyzer to the amplitude of the current marker.
If Normal marker is selected, the reference level will be set to the amplitude
of the current marker.
If Delta, Delta Pair, or Span Pair marker is selected, the reference level will
be set to the amplitude of the Delta Marker.
MkrΔ->CF
Set the center frequency of the analyzer to the frequency difference between the
two markers in Delta, Delta Pair, or Span Pair marker type.
If Normal marker is selected, this function is invalid.
The function is invalid in Zero span mode.
MkrΔ->Span
Set the span of the analyzer to the frequency difference between the two markers
in Delta, Delta Pair or Span Pair marker type.
If Normal marker is selected, this function is invalid.
The function is invalid in Zero span mode.