RIGOL Chapter 4 Troubleshooting&Message
User’s Guide for DSA700 Series
4-18
327 Other language except English not supported.
Any other languages are disabled except English.
328 File operation failed.
329 No space to save file.
The file or folder can not be created or saved owing to limited space.
331 Invalid license key.
The length of the license key must be no more than 20 characters.
332 Invalid File Type to load.
333 Invalid installation, please insert U disk again.
334 Invalid sweep data, confirm to continue?
340 DHCP renew failed.
Fail to configure IP address with DHCP server. Please try manual IP.
341 IP conflict.
The IP address received is already in use. Please try another IP address.
342 Invalid IP.
The IP address received is invalid.
350
Can’t auto-couple sweep time in zero span.
Sweep time can not be coupled automatically in zero span. Please set it
manually.
351
Zero span unavailable for other measurements except T-power.
354 Pre-amplifier can not be enabled.
Please adjust the settings of the reference level, attenuator as well as
maximum mixing level.
355 Scale/div invalid in linear scale mode.
356 Invalid function in zero span.
In zero span, the following operations cannot be executed: Signal Track
On, Span Zoom In, Span Zoom Out, Peak->CF, Mkr->CF, Mkr->Step,
Mkr->Start, Mkr->Stop, Mkr Delta->CF, Mkr Delta->Span, setting the
Mark readout as FREQ or PERIOD.
357 Invalid function in non-zero span.
Video trigger and the1/ΔTime in marker readout are invalid in non-zero
span.
358 Invalid mark readout type.
For Delta Pair marker, the 1/ΔTime marker readout type is invalid; and
for normal marker, MkrΔ->CF and MkrΔ->Span are invalid.
359 Invalid function when trace is not enabled.
A marker could not be assigned to a trace when the trace is not enabled.