360 User preset failed.
The user preset has failed because the edition of the saved status data is
incorrect or the status data has been damaged. In this situation, the
system will execute Factory settings as the defaults.
400 Input signal power out of range.
Indicate that the input signal power is out of range. Working in this state
for a long time would cause damage to the device. Please reduce the
input power.
401
1st LO unlock.
The first LO is unlocked. Please report the error to your RIGOL sales or
technical support.
402 2nd LO unlock.
The second LO is unlocked. Please report the error to your RIGOL sales
or technical support.
412 DA overrange due to volume adjustment in FM.
413 Intermediate frequency signal out of range.
420 Option not installed.
The desired operation cannot be performed because the option required
is not installed.
460 No peak found.
No signal peak that meets the criteria was found under Peak Search.
461 No fundamental wave found.
Indicate that no fundamental wave greater than 50dBm is found.
462 No two-tone signal found.
463 Invalid function as cursor is not enabled.
The cursor cannot be set to the reference level unless it is enabled.
464 Menus under Storage can not be set to Userkey.
465 Invalid function.
470 Calibration memory lost.
Indicate that the calibration data before delivery or last valid calibration
data has been lost. Please report the error to your RIGOL sales or
technical support team.
472
USB firmwarte lost.
473
Device information lost.
474
Data of menu, help, message and font lost.
475
Preset data to load is damaged.