User manual

RIGOL Chapter 5 Specifications
User’s Guide for DSA700 Series
5-4
Displayed Average Noise Level (DANL)
DSA705 DSA710
Frequency
attenuation = 0 dB, RBW = VBW = 100 Hz, sample detector, trace
average ≥ 50, 20 to 30, input impendence = 50 Ω
PA
off
100 kHz to 1 MHz <-90 dBm, <-110 dBm (typ.) <-90 dBm, <-110 dBm (typ.)
1 MHz to 500 MHz <-100 dBm, <-110 dBm (typ.)
<-100 dBm, <-110 dBm (typ.)
500 MHz to 1 GHz
PA
on
100 kHz to 1 MHz <-110 dBm, <-130 dBm (typ.) <-110 dBm, <-130 dBm (typ.)
1 MHz to 500 MHz <-120 dBm, <-130 dBm (typ.)
<-120 dBm, <-130 dBm (typ.)
500 MHz to 1 GHz
Level Display
Logarithmic level axis 1 dB to 200 dB
Linear level axis 0 to reference level
Number of display points 601
Number of traces 3 + math trace
Trace detectors
normal, positive-peak, negative-peak, sample, RMS, voltage
average
quasi-peak (with EMI-DSA800 option)
Trace functions clear write, max hold, min hold, average, view, blank
Units of level axis dBm, dBmV, dBμV, nV, μV, mV, V, nW, μW, mW, W
Frequency Response
DSA705 DSA710
Frequency response
f
c
100 kHz, attenuation = 10 dB, relative to 50 MHz, 20 to
30
PA
off
100 kHz to 500 MHz <0.7 dB
<0.7 dB
500 MHz to 1 GHz
f
c
1MHz, attenuation = 10 dB, relative to 50 MHz, 20 to 30
PA
on
100 kHz to 500 MHz <1.0 dB
<1.0 dB
500 MHz to 1 GHz
Input Attenuation Switching Uncertainty
DSA705 DSA710
Setting range 0 dB to 30 dB, in 1 dB step
Switching uncertainty
f
c
= 50 MHz, relative to 10 dB, 20 to 30
<0.5 dB