User manual

RIGOL Chapter 2 Front Panel Operation
User’s Guide for DSA700 Series
2-14
(dBm) in Log scale type when the scale type is changed from Lin to Log.
In Lin scale type, the Y-axis denotes the linear coordinate; the values shown
at the top of the grid and the bottom of the grid are the reference level and 0
V respectively. Each grid represents 10% of the reference level and the scale
setting function is invalid. The unit of Y-axis will automatically switch to the
default unit (Volts) in Lin scale type when the scale type is changed from Log
to Lin.
The scale type does not affect the unit of Y-axis.
Units
Set the unit of the Y-axis to dBm, dBmV, dBuV, Volts or Watts. Wherein, dBm,
dBmV and dBuV are for Log scale; Volts and Watts are for Linear scale. The default
is dBm.
Key Points:
The conversion relationships between units are as follows.
×=
WR
Volts
B
001.0
1
log10md
2
(2-4)
=
V
Volts
VB
µ
µ
1
log20d
(2-5)
=
V
Volts
VB
m1
log20md
(2-6)
R
Volts
Watts
2
=
(2-7)
Wherein,
R