RIGOL DSA800 Series Specifications DSA800/DSA800E Series Spectrum Analyzer Frequency Frequency DSA815 Frequency range 9 kHz to 1.5 GHz Frequency resolution 1 Hz DSA832E 9 kHz to 3.2 GHz DSA832 9 kHz to 3.2 GHz DSA875 9 kHz to 7.
RIGOL DSA800 Series Specifications Carrier offset DSA815 DSA832E DSA832 DSA875 10 kHz <-80 dBc/Hz <-90 dBc/Hz <-98 dBc/Hz 100 kHz <-100 dBc/Hz (typ.) <-100 dBc/Hz (typ.) <-100 dBc/Hz (typ.) Residual FM 20℃ to 30℃, RBW = VBW = 1 kHz Residual FM DSA815 <50 Hz (nom.) DSA832E DSA832 <20 Hz (nom.
RIGOL DSA800 Series Specifications Resolution bandwidth (-6 dB) (EMI-DSA800 option) 200 Hz, 9 kHz, 120 kHz Amplitude Measurement Range Range fc ≥ 10 MHz DANL to +20 dBm Maximum Input Level DC voltage CW RF power Max.
RIGOL DSA800 Series Specifications NOTE: [1] When fc ≥ 10 MHz, input level > +25 dBm and PA is Off, the protection switch will be on. Displayed Average Noise Level (DANL) DSA815 Frequency attenuation = 0 dB, RBW = VBW = 100 Hz, sample detector, trace average ≥ 50, tracking generator off, 20℃ to 30℃, input impendence = 50 Ω PA off PA on 100 kHz to 1 MHz <-90 dBm, <-110 dBm (typ.) 1 MHz to 1.5 GHz <-110 dBm + 6 × (f/1 GHz) dB, <-115 dBm (typ.) 100 kHz to 1 MHz <-110 dBm, <-130 dBm (typ.
RIGOL DSA800 Series Specifications Displayed Average Noise Level (DANL) (Normalized to 1Hz) DSA815 Frequency DSA832 attenuation = 0 dB, RBW = VBW = 100 Hz, sample detector, trace average ≥ 50, tracking generator off, normalized to 1Hz, 20℃ to 30℃, input impendence = 50 Ω 9 kHz to 100 kHz 100 kHz to 1 MHz 1 MHz to 5 MHz PA off 5 MHz to 1.5 GHz <-110 dBm, <-130 dBm (typ.) <-130 dBm + 6 × (f/1 GHz) dB, <-135 dBm (typ.) 1.5 GHz to 3.2 GHz <-120 dBm (typ.) <-120 dBm (typ.) <-135 dBm, <-138 dBm (typ.
RIGOL DSA800 Series Specifications Displayed Average Noise Level (DANL) DSA832E attenuation = 0 dB, RBW = VBW = 10 Hz, sample detector, trace average ≥ 50, tracking generator off, 20℃ to 30℃, input impendence = 50 Ω PA off PA on Level Display Logarithmic level axis Linear level axis Number of display points Number of traces Trace detectors Trace functions Units of level axis 9 kHz to 100 kHz <-110 dBm (typ.) 100 kHz to 5 MHz <-122 dBm, <-125 dBm (typ.) 5 MHz to 3.2 GHz <-127 dBm, <-130 dBm (typ.
RIGOL DSA800 Series Specifications Frequency Response Frequency response 100 kHz to 1.5 GHz PA off 1.5 GHz to 3.2 GHz 3.2 GHz to 7.5 GHz 100 kHz to 1.5 GHz PA on 1.5 GHz to 3.2 GHz DSA815 DSA832E DSA832 DSA875 fc ≥ 100 kHz, attenuation = 10 dB, relative to 50 MHz, 20℃ to 30℃ <0.7 dB <0.7 dB, <0.5 dB, <0.3 dB (typ.) <0.5 dB, <0.3 dB (typ.) <0.7 dB, <0.3 dB (typ.) fc ≥ 1 MHz, attenuation = 10 dB, relative to 50 MHz, 20℃ to 30℃ <1.0 dB <1.0 dB, 3.2 GHz to 7.5 GHz <0.7 dB, <0.3 dB (typ.) <0.
RIGOL DSA800 Series Specifications Input Attenuation Switching Uncertainty DSA815 Setting range DSA832E DSA832 DSA875 0 to 30 dB, in 1 dB step fc = 50 MHz, relative to 10 dB, 20℃ to 30℃ Switching uncertainty < 0.5 dB < 0.3 dB Absolute Amplitude Uncertainty DSA815 DSA832E DSA832 DSA875 fc = 50 MHz, peak detector, preamplifier off, attenuation = 10 dB, input signal level = -10 dBm, 20℃ to 30℃ Uncertainty <0.4 dB <0.3 dB RBW Switching Uncertainty relative to 1 kHz RBW Uncertainty <0.
RIGOL DSA800 Series Specifications Level Measurement Uncertainty DSA815 DSA832E DSA832 DSA875 95% confidence level, S/N > 20 dB, RBW = VBW = 1 kHz, preamplifier off, attenuation = 10 dB, -50 dBm < input level ≤ 0 dBm, fc > 10 MHz, 20℃ to 30℃ Level measurement uncertainty <1.5 dB (nom.) <1.0 dB (nom.) Page 9/17 <0.8 dB (nom.
RIGOL DSA800 Series Specifications RF Input VSWR DSA815 DSA832E attenuation ≥ 10 dB 300 kHz to 1.5 GHz VSWR 1.5 GHz to 3.2 GHz DSA832 DSA875 <1.5 (nom.) <1.5 (nom.) <1.5 (nom.) <1.5 (nom.) 3.2 GHz to 7.5 GHz <1.8 (nom.
RIGOL DSA800 Series Specifications Distortion Second Harmonic Intercept DSA815 DSA832E DSA832 DSA875 fc ≥ 50 MHz, input signal level = -20 dBm, attenuation = 10 dB Second harmonic intercept (SHI) +40 dBm +40 dBm +45 dBm Third-order Intercept Third-order intercept (TOI) DSA815 DSA832E DSA832 DSA875 fc ≥ 50 MHz, two -20 dBm tones at input mixer spaced by 200 kHz, attenuation = 10 dB +10 dBm +7 dBm +11 dBm, +15 dBm (typ.
RIGOL DSA800 Series Specifications Spurious Response Spurious response DSA815 DSA832E DSA832 DSA875 input terminated 50 Ω, attenuation = 0 dB, 20℃ to 30℃ <-88 dBm (typ.) Intermediate frequency System related sidebands Input related spurious <-90 dBm[1], <-100 dBm (typ.
RIGOL DSA800 Series Specifications Trigger Trigger Trigger source External trigger level free run, video, external 5 V TTL level SSC-DSA (Option) (Only for DSA815) Signal Seamless Capture (SSC) Measurement bandwidth 1.
RIGOL DSA800 Series Specifications Input /Output Front Panel Connectors impedance RF input connector Tracking generator impedance output connector 50 Ω (nom.) N female 50 Ω (nom.) N female Internal/ External Reference Internal reference External reference frequency output level 10 MHz +3 dBm to +10 dBm, +8 dBm (typ.) impedance connector frequency input level impedance connector 50 Ω (nom.) BNC female 10 MHz ± 5 ppm 0 dBm to +10 dBm 50 Ω (nom.
RIGOL DSA800 Series Specifications Mass Memory Mass memory flash disk (internal), USB storage device (not supplied) Power Supply Input voltage range, AC AC supply frequency Power consumption 100 V to 240 V (nom.) 45 Hz to 440 Hz 35 W (typ.), max. 50 W with all options Environmental Temperature Humidity Altitude operating temperature range storage temperature range 0℃ to 30℃ 0℃ to 50℃ -20℃ to 70℃ ≤ 95% rel. humidity 30℃ to 40℃ ≤ 75% rel.
RIGOL DSA800 Series Specifications Ordering Information Description Order Number spectrum analyzer, 9 kHz to 1.5 GHz (with preamplifier) DSA815 spectrum analyzer, 9 kHz to 3.2 GHz DSA832E spectrum analyzer, 9 kHz to 3.2 GHz DSA832 spectrum analyzer, 9 kHz to 7.5 GHz DSA875 spectrum analyzer, 9 kHz to 1.5 GHz (with preamplifier, with tracking Model generator, factory installed) spectrum analyzer, 9 kHz to 3.2 GHz (with tracking generator, factory installed) spectrum analyzer, 9 kHz to 3.
RIGOL DSA800 Series Specifications Page 17/17