Instructions

Chapter 2 Front Panel Operation RIGOL
User’s Guide for DSA800 Series
2-23
Sweep and Function Settings
BW/Det
Set the RBW (Resolution Bandwidth), VBW (Video Bandwidth) and detector type
parameters of the analyzer.
RBW
Set the desired resolution bandwidth in order to distinguish between signals which
are close in frequency.
Key Points:
Reducing RBW will increase the frequency resolution, but will increase the
sweep time (Sweep Time is affected by a combination of RBW and VBW when
it is in Auto mode).
RBW varies with the span (non-zero span) in Auto RBW mode.
You can modify this parameter using the numeric keys, knob or direction keys.
For more details, please refer toParameter Setting.
Note: If the detector type is Quasi-Peakor the filter type is EMI, RBW can be
200 Hz, 9 kHz or 120 kHz only.
Table 2-13 RBW (Filter Type is Gauss)
Parameter Explanation
Default 1 MHz
Range 10 Hz to 1 MHz
Unit GHz, MHz, kHz, Hz
Knob Step in 1, 3, 10 sequence
Direction Key Step in 1, 3, 10 sequence