Specifications

RIGOL
DS6000 Users Guide
5-11
When the Pulse Condition is set to , , or , press
Set Up and use to input the desired value. The range available is from 4 ns
to 4 s.
When the Pulse Condition is set to or , press Upper Limit and
Lower Limit respectively and use to input the desired values. The range of
the upper limit available is from 12 ns to 4 s. The range of the lower limit
available is from 4 ns to 3.99 s. Note that the lower limit of the pulse width must
be lower than the upper limit.
Trigger Mode:
Press Sweep to select the Trigger Mode (page 5-3) under this trigger type as auto,
normal or single. The corresponding status light of the current trigger mode turns on.
Trigger Setting:
Press Set Up to set the trigger parameters (trigger coupling, trigger holdoff and
noise rejection) under this trigger type.
Trigger Level:
Use TRIGGER LEVEL to modify the level. For details, please refer to the
description of Trigger Level on page 5-9.
Trigger level
Positive
pulse width
A
B