Specifications

RIGOL
DS6000 Users Guide
5-13
When the Slope Condition is set to , , or ,
press Time and use to input the desired value. The range available is from
10 ns to 1 s.
When the Slope Condition is set to or , press Upper Limit
and Lower Limit respectively and use to input the desired values. The
range of time upper limit available is from 20 ns to 1 s. The range of the time
lower limit available is from 10 ns to 1 s. Note that the time lower limit must be
lower than the upper limit.
Vertical Window:
Press Vertical to select the desired vertical window.
You can select the boundaries of the trigger level and then use TRIGGER LEVEL
to adjust the trigger level. During the adjustment, two orange trigger level lines and
two trigger marks ( and ) appear on the screen and move up and down with
the turning of the knob, while at the same time, the trigger level value and the slope
value are displayed at the lower left corner of the screen. When stopping turning the
knob, the trigger level lines and trigger marks disappear in about 2 seconds.
Slew Rate= | (V
A
-V
B
) / (T
A
-T
B
) |
The adjustment mode of the trigger level is different when different vertical window
is selected:
: only adjust the upper limit of the trigger level. During the adjustment, the
“UP Lvl” and the “Slew Rate” change accordingly but the “Low Lvl” remains
unchanged.
: only adjust the lower limit of the trigger level. During the adjustment, the
“Low Lvl” and the “Slew Rate” change accordingly but the “UP Lvl” remains
The upper limit of trigger level (Up Lvl)
The lower limit of trigger level (Low Lvl)
A
B
Positive
Slope time