Specifications

RIGOL
DS6000 Users Guide
5-17
Pattern Trigger
Identify the trigger condition through searching for the specified pattern. Pattern is
the logic AND combination of each channel and the value of each channel can be H
(High), L (Low) or X (Dont Care). One channel in the pattern can be specified with
rising edge or falling edge and if the pattern settings of all the other channels are
True (H or L), the oscilloscope will trigger on the specified edge. If no edge is
specified, the oscilloscope will trigger on the last edge that makes the pattern as
True. If patterns of all the channels are set to Dont Care, the oscilloscope would
not trigger.
Trigger Type:
Press Type to select Pattern. At this point, the trigger setting information as shown
in the figure below is displayed at the upper right corner of the screen.
Channel Selection:
Press Source to select CH1-CH4 or EXT as the Trigger Source for H, L, X or edge
condition. The current trigger source is displayed at the upper right corner of the
screen.
Pattern Setting:
Press Code to set the pattern code of the current channel. At this point, the top of
the screen will show the pattern setting area, see below.
: set the pattern of the channel selected to H, namely the voltage level is
higher than the threshold level of the channel.
: set the pattern of the channel selected to L, namely the voltage level is
lower than the threshold level of the channel.
: set the pattern of the selected channel to Dont Care, namely this channel is
not a part of the pattern. When all the channels in the pattern are set to Dont
Care, the oscilloscope will not trigger.
or : set the pattern to the rising or falling edge of the channel selected.
Note that in the pattern, you can only specify one rising edge or falling edge. If one
edge item is currently defined and then another channel in the pattern is selected to