CAN Trigger
Trigger on the specified frame type of the data frame. When using CAN trigger, you
need to specify the signal source, signal rate and trigger signal type of the CAN
signal.
Trigger Type:
Press Type to select CAN. At this point, the trigger setting information as shown in
the figure below is displayed at the upper right corner of the screen.
Source Selection:
Press Source to select CH1-CH4, EXT or EXT/5 as the Trigger Source. The current
trigger source is displayed at the upper right corner of the screen.
Signal Type:
Press Signal Type to select the desired signal type.
Rx: receiving signal from the CAN signal line.
Tx: transmission signal from the CAN signal line.
CAN_H: actual CAN_H differential bus signal.
CAN_L: actual CAN_L differential bus signal.
Differential: CAN differential bus signal connected to the analog channel by
using the differential probe.
Trigger Condition:
Press When to select the desired trigger condition.
SOF: trigger on the start frame position of the data frame.
EOF: trigger on the end frame position of the data frame.
Frame Type: after this type is selected, press Frame Type to select to trigger
on “Data”, “Remote”, “Error” or “OverLoad”. The following parameters are
available for other frame types except OverLoad:
“Data” trigger:
-- Press the ID Setup softkey and select Specific IDor Random ID. When
Specific ID” is selected, press the ID Format softkey and selectStandardor
Expand”; press the ID Data softkey and use to set the desired value. The
range is from 0 to 2047 (standard ID format) or 0 to 536870911 (expanded ID