format).
-- Press the Bits softkey and use to set the desired value. The range is from
1 to 8.
-- Press the CurrentBit softkey to select the desired data bit and the range is
from 0 to (Bits×8-1).
--Press the Data softkey to set the data pattern of the current bit to X, H or L.
--Press the AllBits softkey to set the data pattern of all the data bits to the data
pattern specified in Data.
“Remote” trigger:
--Press the ID Setup softkey and select “Specific ID” or “Random ID”. When
“Specified ID” is selected, press the ID Format softkey and select “Standard”
or “Expand”; press the ID Data softkey and use to set the desired value.
The range is from 0 to 2047 (standard ID format) or 0 to 536870911 (expanded
ID format).
“Error” trigger:
Press the Error Type softkey to select “Bit Fill”, “Answer Error”, “Check Error”,
“Format Error” or “Random Error”.
Baud:
Press Baud Rate to set the CAN baud to match the CAN bus signal and the values
available are 10 kb/s (default), 20 kb/s, 33.3 kb/s, 50 kb/s, 62.5 kb/s, 83.3 kb/s, 100
kb/s, 125 kb/s, 250 kb/s, 500 kb/s, 800 kb/s, 1 Mb/s or user. After “User” is selected,
press User and use to input the desired rate within 1 kb/s and 10.0 Mb/s.
Sample Point:
Sample point is the point within the bit’s time. The oscilloscope samples the bit level
at this point. “Sample point” is represented by the percentage of “the time from the
start of the bit’s time to the sample point” in the “bit’s time”.
1Bit
60%
70%
80%