Specifications

RIGOL
DS6000 Users Guide
10-8
5. Previous
After the waveform analysis finishes, press the Previous softkey can locate the
error frame before the current error frame. Pressing the Run/Stop button can
also perform the operation.
6. Next
After the waveform analysis finishes, press the Next softkey can locate the
error frame following the current error frame. Pressing the Single button can
also perform the operation.
7. Current Frame
Press the Current Frame softkey to adjust the data frame currently displayed
and the adjustable range is from 1 to the total number of frames in this
waveform analysis. Using the navigation knob can also perform the operation.
8. Setup
Press the Setup softkey to open the detailed setting menu.
Screen Start: press this key to set the start point of waveform analysis and
the range is from 5 to 695. The start point must be less than the “Screen
End” currently set.
Screen End: press this key to set the end point of waveform analysis and
the range is from 1 to 699. The end point must be greater than the “Screen
Start” currently set.
Start Frame: press this key to set the start frame of waveform analysis and
the default is the first frame.
End Frame: press this key to set the end frame of waveform analysis and
the default is the last frame.
Threshold: press this key to set the threshold of waveform analysis and the
range is from 1% to 99%. The threshold is used to judge whether the frame
is an error frame. A frame can be recognize as an error one if the (relative)
difference value between the frame and the templet is equal to or larger
than the threshold currently set.
9. TempletDisp
Press the TempletDisp softkey to enable or disable templet display. For the
templet setting, refer to the introductions in Analysis Based on Trace and
Analysis Based on Pass/Fail Mask.