Specifications

RIGOL
DS6000 Users Guide
10-10
Analysis Based on Pass/Fail Mask
Press the Analyze softkey and select “Pass/Fail”. Then, set the templet used in
analysis based on Pass/Fail mask through the method below.
1. X Mask
Press this softkey and use to adjust the horizontal threshold and the range
is from 0.02 div to 4.00div. During the adjustment, two curves would be
displayed to show the outline of current mask, see the figure below.
2. Y Mask
Press this softkey and use to adjust the vertical threshold and the range is
from 0.03 div to 4.00 div. During the adjustment, two curves would be displayed
to show the outline of current mask.
3. Create Mask
Mask is the templet used in waveform analysis. Press the Create Mask softkey
to immediately apply the mask (X Mask and Y Mask) currently created.
4. Save
Press the Save softkey to enter the file storage interface. Store the test mask
file to internal or external memory by referring to the introduction in Store and
Recall.
5. Load
Press the Load softkey to enter the file load interface. Load the test mask into
the internal memory of the instrument by referring to the introduction in Store
and Recall.
Current Mask Outline