Specifications

RIGOL
DS6000 Users Guide
4-8
Memory Depth
Memory depth refers to the number of waveform points that the oscilloscope can
store in a single trigger sample and it reflects the storage ability of the acquisition
memory. DS6000 provides a memory depth up to 140 M points.
T
Trigger Point
Pre-sample Delayed Sample
Memory Depth
The relationship of memory depth, sample rate and waveform length fulfills the
following equation:
Memory depth = sample rate (Sa/s) × waveform length (s/div × div)
Press ACQ Mem Dept, use to select the desired memory depth (the default
is auto) and then press the knob to select the option. You can also press the Mem
Dept softkey continuously to switch the memory depth.
When single channel is turned on, the memory depths available include: Auto,
14kPoints, 140kPoints, 1.4MPoints, 14MPoints and 140MPoints. In Auto mode, the
oscilloscope selects the memory depth automatically according to the current sample
rate.
When double channels are turned on (CH1 and CH2 or CH3 and CH4), the memory
depths available include: Auto, 7kPoints, 70kPoints, 700kPoints, 7MPoints and
70MPoints. In Auto” mode, the oscilloscope selects the memory depth automatically
according to the current sample rate.