User`s guide

RIGOL Chapter 4 Application Examples
4-6 DSG3000 Users Guide
Pulse Train Application
Output a user-defined pulse train from [PULSE IN/OUT] connector. The specific
parameters to be set in the pulse list as shown in the table below.
SN
On Time
Off Time
Repeat
1
10ms
30ms
2
2
20ms
40ms
1
1. Restore to factory setting
Press System Reset Preset Type “Factory” and then press Preset to
restore the factory setting.
2. Install Pulse Train Generator (Option)
Please refer to
DSG3000 Firmware and Options Installation Instruction
to install
and activate the pulse train generator (option PUG-DSG3000).
3. Edit the Pulse Train List
Press Pulse to use the page up/down key
to open the 3/3 menu page,
and then press Train List to enter the editing interface of pulse train list.
Insert
If the parameter values in the current list are the only one row, press
Insert to insert a row with the same pulse value below the current row.
Delete
If the parameter values in the current list are more than two lines, you can
press Delete to delete the pulse value of the row currently selected, and
last retain the only two rows parameters.
Edit
Use the direction keys or knob to move the cursor and select the
parameters of the first row to be edited.
On Time: use the numeric keyboard to input the value 10, and select
the desired unit (ms) from the pop-up unit menu or unit keys.
Off Time: use the numeric keyboard to input the value 30, and select
the desired unit (ms) from the pop-up unit menu or unit keys.
Repeat: use the numeric keyboard to input the value 2, and then press
Enter softkey or Enter.
Use the same method to set the pulse parameters of the second row.
NOTE: Dwell time is calculated automatically by the three parameters (On
Time, Off Time and Repeat) to be set.