User`s guide
RIGOL Chapter 6 Specifications
6-4 DSG3000 User’s Guide
Level
Setting Range
Specification level range
Setting range
Maximum output
level
9kHz ≤ f <100kHz
+7dBm
+10dBm
100kHz ≤ f <1MHz
+13dBm
+15dBm
1MHz ≤f ≤3GHz
+13dBm
+25dBm
3GHz <f ≤6GHz
+13dBm
+20dBm
Minimum output
level
9kHz ≤ f <100kHz
-110dBm
-120dBm
100kHz ≤ f ≤ 6GHz
-130dBm
-140dBm
Setting resolution
0.01dB
Absolute Level Uncertainty
[1]
Level
uncertainty
+13 to
-60dBm
-60 to
-110dBm
-110 to -130dBm
9kHz ≤f<100kHz
≤ 0.5dB (typ.)
≤ 0.7dB (typ.)
100kHz ≤f≤3GHz
≤ 0.7dB,
≤ 0.5 (typ.)
≤ 0.9dB,
≤ 0.5 (typ.)
≤ 0.7dB (typ.)
3GHz < f ≤ 6GHz
≤ 0.9dB,
≤ 0.5 (typ.)
≤ 1.1dB,
≤ 0.5 (typ.)
≤ 0.9dB (typ.)
VSWR
[2]
1MHz ≤ f ≤ 6GHz
< 1.8 (typ.)
Level Setting
Setting time
ALC state on, frequency fixed,
temperature range: 20℃ to
30℃
≤ 5ms (typ.)
Uninterrupted level
setting range
ATT fixed mode, ALC state on,
level range -110dBm to
+13dBm
>20dB (typ.)
Max. Reverse Power
Max. reverse power
Max. DC voltage
50V
1MHz < f ≤ 6GHz
10W
NOTE: [1] ALC state: on or auto mode, 20℃ to 30℃
[2] In 50Ω system, typical, level ≤-10dBm, ATT auto mode