User`s guide

RIGOL Chapter 2 Front Panel Operations
2-8 DSG3000 Users Guide
1. Set the flatness calibration state
Press LEVEL and use the page up/down key
to open the 2/2 menu page;
then, press Flatness Switch to set the flatness calibration to Onor Off.
When the flatness calibration is enabled, the UF label will be displayed in the
amplitude area in the user interface.
2. Create the flatness calibration list
Press LEVEL Flatness Cal List to enter the editing interface of flatness
calibration list.
SN
Row number. You can use Insert and Delete to add and delete
rows.
Freq
Frequency. The flatness calibration list can contain the amplitude
calibration values of up to 6000 frequency points. The 6000
frequency points can locate in any frequency segment within the
maximum frequency range available.
You can save calibration lists applicable to different test settings or
frequency ranges and recall different calibration list when required.
Level
Amplitude calibration value. When calibrating the amplitude of the
current frequency point, only the actual output amplitude will be
affected and the amplitude display value will not change.
Insert
Press Insert to insert a row with the same calibration value below the
current row. At this point, the number of rows of the flatness calibration list
increases by one.
Delete
Press Delete to delete the calibration value of the row currently selected.
At this point, the number of rows of the flatness calibration list reduces by
one.
Goto Row
Press Goto Row to select the desired row for editing in the pop-up menu.
It is usually used to modify the calibration value of the specified row.
N: press n and use the numeric keyboard to input the desired row number
to select that row. The row number input should not be greater than the
total number of rows of the current list.
Top: press Top to select the first row of the current list.
Mid: press Mid to select the row at the middle of the current list.
Bottom: press Bottom to select the last row of the current list.
When the desired row is selected, press
to return to the previous
menu to modify the calibration value of the row currently selected.