Chapter 2 Front Panel Operations RIGOL
DSG3000 Users Guide 2-17
*TRGcommand is sent.
External Trigger
In external trigger, the RF signal generator accepts the trigger signal input
from the [TRRIGER IN] connector (as shown in the figure below) at the
rear panel. If the sweep mode is set to Cont, the instrument starts a
sweep each time a TTL pulse signal with specified polarity is received. If the
sweep mode is Single, press Single to meet the single sweep condition.
After that the instrument starts a sweep and then stops each time a TTL
pulse signal with the specified polarity is received.
To specify the polarity of the TTL pulse signal, press Trig Slope to select
Posor Negand the default is Pos.
2. Point Trigger Mode
Select the trigger mode of each sweep point in a sweep period.
Press SWEEP and use the page up/down key
to open the 2/3 menu page;
then, press Point Trig to select Auto, Key, Busor Exttrigger.
NOTE: The following descriptions are valid when the corresponding trigger
mode of sweep period is met.
Auto Trigger
The default mode is auto. If the sweep mode is set to Cont, the
instrument will start sweeping each sweep point continuously within a
sweep period once a sweep type is selected. If the sweep mode is set to
Single”, press Single to meet the single sweep condition. After that the
instrument starts to sweep and then stops after the sweep period is
finished.
Key Trigger
After selecting Keytrigger, if the sweep mode is set to Cont, the
instrument starts to sweep a point each time Trigger at the front panel is
pressed. If the sweep mode is Single, press Single to meet the single
sweep condition. At this point, the instrument starts to sweep a point and
stops after the sweep period is finished each time Trigger at the front
panel is pressed.