User`s guide

RIGOL Chapter 2 Front Panel Operations
2-28 DSG3000 Users Guide
NOTE: For Trainpulse width and the related parameter settings, please refer to
Pulse Train List.
1. Single Pulse
Pulse Width
Period
Press Width to set the width of the single-pulse modulating signal.
Use the numeric keyboard or knob to input the desired pulse width value.
The range of the pulse width is from 10 ns to 170 s10 ns.
The pulse width is limited by the minimum pulse width and pulse period.
Pulse Width Minimum Pulse Width
Pulse Width Pulse Period10 ns
2. Double Pulse
#1 Pulse Width
Period
Delay
#2 Pulse Width
Press Width to set the #1 pulse width of the double-pulse modulating signal.
Use the numeric keyboard or knob to input the desired pulse width value.
The range of the #1 pulse width is from 10 ns to 170 s 10 ns.
The #1 pulse width is limited by the minimum pulse width and delay.
Pulse Width Minimum Pulse Width
Pulse Width Delay 10 ns
Press #2 Width to set the #2 pulse width of the double-pulse modulating
signal.
Use the numeric keyboard or knob to input the desired pulse width value.
The range of the #2 pulse width is from 10 ns to 170 s – 10 ns.
The #2 pulse width is limited by the minimum pulse width, pulse period and
delay.
#2 Pulse Width #2 Minimum Pulse Width
#2 Pulse Width Pulse Period Delay 10 ns