Chapter 2 Front Panel Operations RIGOL
DSG3000 Users Guide 2-41
Explanation
The ranges of sine and sweep-sine frequency are from 100 mHz to 1 MHz.
The range of square waveform frequency is from 100 mHz to 20 kHz.
The ranges of triangle and ramp waveform frequency are from 100 mHz to
100 kHz.
Sweep-Sine Waveform
The sweep-sine waveform refers to sine waveform changes gradually from the start
frequency to stop frequency within the specified time, namely frequency sweep.
1. Select the sweep-sine
Press LF Waveform to selectSwp-Sine. At this point, the RF
signal generator output sweep-sine signal from [LF OUTPUT] connector at the
front panel according to the current setting.
2. Set the sweep-sine parameters
Start Frequency
Press Start Freq, use the numeric keyboard to input the value of the start
frequency and select the desired unit from the pop-up unit menu or unit
keys. You can also press Enter to select the default unit (Hz).
Stop Frequency
Press Stop Freq, use the numeric keyboard to input the value of the stop
frequency and select the desired unit from the pop-up unit menu or unit
keys. You can also press Enter to select the default unit (Hz).
NOTE: After the Start Freqor Stop Freqis modified, the instrument will
restart the sweep and output from the Start Freq to Stop Freq.
Amplitude
Press Level, use the numeric keyboard to input the value of the amplitude
and select the desired unit from the pop-up unit menu or unit keys. You can
also press Enter to select the default unit (V).
Sweep Time
The sweep time refers to the duration of a sweep.
Press Sweep Time, use the numeric keyboard to input the time value and
select the desired unit from the pop-up unit menu or unit keys. You can also
press Enter to select the default unit (s).