RIGOL Chapter 2 Front Panel Operations
2-46 DSG3000 Users Guide
Average Measurement
Average Mode
Press AVG MOD to select Manulor Autoaverage measurement mode.
Manul: in this mode, you can set theAVG NUMof the measurement value
manually.
Auto: the instrument configures the AVG NUM according to the measurement
value of the current signal automatically.
Average Number
Press AVG NUM to set the average number of the signal average measurement
manually. The range is from 1 to 1024.
Tips
1. When the Autoaverage mode is selected, the average mode will switch
to the Manual mode
automatically after you set the average number via
AVG NUM.
2. Increasing the number of averages can reduce the measurement noise,
but will increase the measurement time.
Preset
Press Preset and the power meter parameters will be restored to the factory default
settings.
Limit Measurement
In limit measurement, the measurement is performed within the specified limit range.
If the measurement value exceeds this range, the measurement fails.
Measurement Type
Press Type to select the limit measurement type, including Upper, Lower, Both
or Off.
Upper: select the maximum limit measurement. When the alert function is
turned on and the measurement value exceeds the maximum limit, the
instrument triggers the alarm.