Instructions
RIGOL  Chapter 5 Troubleshooting 
5-4   DSG800 User Guide 
(3)  Check whether the RF signal generator is working under the specified 
environment temperature. 
(4)  Check whether the test is being performed in a highly magnetic 
environment. 
(5)  Check whether the power supplies of the RF signal generator and test 
system have strong interference.  
(6)  Check whether the performance of the test device meets the requirement.  
(7)  Check whether the test device is within the calibration period. 
(8)  Check whether the test device is working under the required conditions of 
the corresponding manual. 
(9)  Check whether all the connections are tight. 
(10) Check whether the cables have internal damage. 
(11) Make sure that the operations conform to the settings and processes 
required by the performance verification guide. 
(12) Check whether the error calculation is correct. 
(13) Get a correct understanding of the definitions of "Typical Value" and 
"Nominal Value". 
—  Typical Value: the performance specifications of this product under the 
specified conditions. 
—  Nominal Value: the approximate quantity of the product during 
application. 