Instructions
RIGOL  Chapter 2 Front Panel Operations 
2-10   DSG800 User Guide 
F/L
T
Start Frequency
/Start Level
Start Frequency
/Start Level
Ramp
Triangle
Stop Frequency
/Stop Level
Stop Frequency
/Stop Level
Ramp: the sweep period always starts from the start frequency or start level 
to the stop frequency or stop level and the sweep sequence is similar to a 
ramp waveform.  
Triangle: the sweep period always starts from the start frequency or start 
level to the stop frequency or stop level and then returns back to the start 
frequency or start level. The sweep sequence is similar to a triangle 
waveform.  
Sweep Mode 
Press SWEEP  Mode to select "Cont" or "Single" sweep and the default is "Cont".  
  Cont: when continuous sweep is selected, the continuous sweep label is 
displayed in the frequency or amplitude area in the user interface. When the 
trigger condition is met, the instrument sweeps continuously according to the 
current setting.  
  Single: when single sweep is selected, the single sweep label is displayed in the 
frequency or amplitude area in the user interface. When the trigger condition is 
met, the instrument performs a single sweep according to the current setting 
and then stops.  
Single Sweep 
If the current sweep mode is "Cont", press Single to switch the sweep mode to 
"Single" and the instrument will perform a single sweep if the trigger condition is 










