Instructions
RIGOL  Chapter 2 Front Panel Operations 
2-50   DSG800 User Guide 
3)  Input the option key, instrument serial number (press Syst  
Information  System Info to acquire the instrument serial number) 
and indentifying code; click "Generate" to acquire the corresponding option 
license.  
Tips: 
You can also install an option by operating the RF signal generator remotely.  
1. 
Build the communication between the RF signal generator and PC. You can 
use the USB or LAN interface. For the connecting methods, please refer to 
"
Remote Control".  
2.  Send the :SYSTem:LKEY <license key> command (for 
example, :SYSTem:LKEY QA7ZCZEH6AC54SFNKA853MS5CB3A); the RF 
signal generator identifies the license received and matches it with the 
corresponding option. At this point, the corresponding option is installed 
and activated.  
Self-test 
1.  Screen Test 
Press Screen Test to enter the screen test interface. The system provides five 
colors (include white, blue, green, red and black) to test whether the display 
contains bad pixels. Press any key to switch the screen color and exit the test.  
2.  Key Test 
Press Key Test to enter the keyboard test interface. Press the function keys at 
the front panel one by one and check whether the corresponding key is 
illuminated; if not, failure might occur to that key. Press ESC three times to exit 
the test. 
Note: If the keys at the front panel are transparent, the corresponding 
backlight will go on during the test. 
Sanitation 
Press Syst  Sanitation to clear all the data set by users and restore them to the 
factory state, including: 
  Format NAND FLASH. 
  Restore the user data saved in the NVRAM and NorFlash to factory settings. 
  Restore the HOST NAME, IP address and password in LXI to factory settings. 