Instructions
Chapter 2 Command System  RIGOL 
DSG800 Programming Guide  2-21 
[:SOURce]:FM[:DEViation]:STEP[:INCRement] 
Syntax  [:SOURce]:FM[:DEViation]:STEP[:INCRement] <value> 
[:SOURce]:FM[:DEViation]:STEP[:INCRement]? 
Description  Set the FM frequency deviation step. 
Query the FM frequency deviation step. 
Parameter 
Name 
Type 
Range 
Default 
<value> 
Real 
10mHz to 500kHz 
1kHz 
Explanation    When <value> is set in "Number" form, the default unit is Hz; for example, 
5000. In addition, <value> can also be set in "Number + Unit" form; for 
example, 5kHz.  
  After the frequency deviation step is set, you can rotate the knob to modify the 
deviation at the current step. At this point, You can query or set the current 
frequency deviation using the [:SOURce]:FM[:DEViation] command. 
Return Format  The query returns the FM frequency deviation step. For example, 5.00000kHz. 
Example 
:FM:STEP:INCR 5kHz 
:FM:STEP:INCR? 
Related 
Command 
[:SOURce]:FM[:DEViation] 
[:SOURce]:FM:EXT:COUP 
Syntax  [:SOURce]:FM:EXT:COUP AC|DC 
[:SOURce]:FM:EXT:COUP? 
Description  Set the coupling mode of FM external modulation. 
Query the coupling mode of FM external modulation. 
Parameter 
Name 
Type 
Range 
Default 
AC|DC  Discrete  AC|DC  AC 
Explanation    AC: set the coupling mode of FM external modulation to "AC". 
  DC: set the coupling mode of FM external modulation to "DC". 
  When the modulation source of FM is set to "Int", this command is invalid. 
Return Format  The query returns AC or DC. 
Example  :FM:EXT:COUP AC 
:FM:EXT:COUP? 
Related 
Command 
[:SOURce]:FM:SOURce 










