Instructions
Chapter 2 Command System  RIGOL 
DSG800 Programming Guide  2-29 
[:SOURce]:IQ:BASeout:LEVel 
Syntax  [:SOURce]:IQ:BASeout:LEVel <value> 
[:SOURce]:IQ:BASeout:LEVel? 
Description  Set the baseband output amplitude. 
Query the baseband output amplitude. 
Parameter 
Name 
Type 
Range 
Default 
<value>  Real  20mV to 1.5V  1V 
Explanation    When <value> is set in "Number" form, the default unit is V. Besides, <value> 
can also be set in "Number + Unit" form; for example, 1.1V.  
  The default unit of the return value is V. 
  After the baseband output amplitude is set, you can rotate the knob to modify 
the amplitude at the current step. At this point, you can set and query the 
current step using the [:SOURce]:IQ:BASeout:LEVel:STEP command. 
Return Format  The query returns the amplitude of the baseband output signal. For example, 
1.100000. 
Example  :IQ:BAS:LEV 1.1 
:IQ:BAS:LEV? 
Related 
Command 
[:SOURce]:IQ:BASeout:LEVel:STEP 
[:SOURce]:IQ:BASeout:LEVel:STEP 
Syntax  [:SOURce]:IQ:BASeout:LEVel:STEP <value> 
[:SOURce]:IQ:BASeout:LEVel:STEP? 
Description  Set the baseband output amplitude step.  
Query the baseband output amplitude step. 
Parameter 
Name  Type  Range  Default 
<value> 
Real 
0.01V to 0.5V 
0.1V 
Explanation    When <value> is set in "Number" form, the default unit is V. Besides, <value> 
can also be set in "Number + Unit" form; for example, 0.2V.   
  The default unit of the return value is V. 
  After the baseband output amplitude step is set, you can rotate the knob to 
modify the baseband output amplitude at the current step. At this point, you can 
query or set the baseband output amplitude using the 
[:SOURce]:IQ:BASeout:LEVel command. 
Return Format  The query returns the baseband output amplitude step. For example, 0.200000. 
Example 
:IQ:BAS:LEV:STEP 0.2 
:IQ:BAS:LEV:STEP? 
Related 
Command 
[:SOURce]:IQ:BASeout:LEVel 