Instructions
RIGOL Chapter 2 Front Panel Operations
2-8 DSG800 User Guide
Step Sweep
1. Select the step sweep mode
Press SWEEP Type to select "Step". At this point, the RF signal generator
performs step sweep according to the current settings.
2. Set the sweep parameters
Press SWEEP and use the menu page up/down key
to open the 2/3 page
of the menu; then, press Step Swp to set the sweep parameters, such as the
start frequency, stop frequency, start level, stop level and number of sweep
points.
Start Frequency
Press Start Freq, use the numeric keyboard to input the start frequency
value and select the desired unit from the pop-up unit menu. You can press
Backspace to delete the number at the left of the cursor.
Stop Frequency
Press Stop Freq, use the numeric keyboard to input the stop frequency
value and select the desired unit from the pop-up unit menu. You can press
Backspace to delete the number at the left of the cursor.
Explanation
The start frequency and stop frequency are the frequency upper and
lower limits of frequency sweep respectively.
When the sweep direction is forward, the RF signal generator
sweeps from the start frequency to the stop frequency.
When the sweep direction is downward, the RF signal generator
sweeps from the stop frequency to the start frequency.
When the "Start Freq" or "Stop Freq" is modified, the RF signal
generator will restart the sweep and output from the specified "Start
Freq" or "Stop Freq".
Start Level
Press Start Lev, use the numeric keyboard to input the start level value
and select the desired unit from the pop-up unit menu. You can press
Backspace to delete the number at the left of the cursor.
Stop Level
Press Stop Lev, use the numeric keyboard to input the stop level value and
select the desired unit from the pop-up unit menu. You can press
Backspace to delete the number at the left of the cursor.