Instructions
Chapter 2 Command System RIGOL
DSG800 Programming Guide 2-39
[:SOURce]:IQ:TRIGger:SEGMent:MODE
Syntax [:SOURce]:IQ:TRIGger:SEGMent:MODE SAME|NEXT|SEAMLESS|SEQUENCER
[:SOURce]:IQ:TRIGger:SEGMent:MODE?
Description Set the trigger mode of the IQ data segment.
Query the trigger mode of the IQ data segment.
Parameter
Name
Type
Range
Default
SAME|NEXT|SEAMLESS
|SEQUENCER
Discrete
SAME|NEXT|SEAMLESS
|SEQUENCER
SAME
Explanation SAME: select the "Same Seg" mode. At this point, the RF signal generator
outputs the same trigger segment.
NEXT: select the "Next Seg" mode. At this point, the instrument will switch to
the output of the next segment when the
[:SOURce]:IQ:TRIGger:SEGMent:EXECute command is sent.
SEAMLESS: select the "Seamless" mode. At this point, the instrument will switch
to the output of the next segment seamlessly when the
[:SOURce]:IQ:TRIGger:SEGMent:EXECute command is sent. This parameter is
valid only when the sample rates of all the waveform segments in the
"Segment" file loaded are the same.
SEQUENCER: select the "Sequencer" mode. At this point, the instrument
outputs each waveform segment according to the "Seg List" (*.SEQ) file or "Seg
csv" (*.CSV) file loaded. At this point, the
[:SOURce]:IQ:TRIGger:SEGMent:NEXT and
[:SOURce]:IQ:TRIGger:SEGMent:EXECute commands are invalid. In addition,
this parameter is valid only when the "Seg List" file or "Seg csv" file is loaded.
The command is valid only when the "Segment" file is loaded.
Return Format
The query returns the trigger mode of the IQ data segment. For example, SAME.
Example :IQ:TRIG:SEGM:MODE SAME
:IQ:TRIG:SEGM:MODE?
Related
Commands
[:SOURce]:IQ:TRIGger:SEGMent:EXECute
[:SOURce]:IQ:TRIGger:SEGMent:NEXT