Instructions
Chapter 2 Command System RIGOL
DSG800 Programming Guide 2-47
[:SOURce]:PM[:DEViation]:STEP[:INCRement]
Syntax [:SOURce]:PM[:DEViation]:STEP[:INCRement] <value>
[:SOURce]:PM[:DEViation]:STEP[:INCRement]?
Description Set the phase deviation step of ØM.
Query the phase deviation step of ØM.
Parameter
Name
Type
Range
Default
<value>
Real
0.01rad to 2.5rad
1rad
Explanation When <value> is set in "Number" form, the default unit is rad. Besides,
<value> can also be set in "Number + Unit" form; for example, 1rad.
The default unit of the return value is rad.
After the phase deviation step is set, you can rotate the knob to modify the
phase deviation at the current step. At this point, you can query or set the
phase deviation using the [:SOURce]:PM[:DEViation] command.
Return Format The query returns the phase deviation step. For example, 1.000000.
Example :PM:DEV:STEP 1
:PM:DEV:STEP?
Related
Command
[:SOURce]:PM[:DEViation]
[:SOURce]:PM:EXT:COUP
Syntax [:SOURce]:PM:EXT:COUP AC|DC
[:SOURce]:PM:EXT:COUP?
Description Set the coupling mode of ØM external modulation.
Query the coupling mode of ØM external modulation.
Parameter
Name Type Range Default
AC|DC
Discrete
AC|DC
AC
Explanation AC: set the coupling mode of ØM external modulation to "AC".
DC: set the coupling mode of ØM external modulation to "DC".
When the modulation source of ØM is set to "Int", this command is invalid.
Return Format The query returns AC or DC.
Example
:PM:EXT:COUP AC
:PM:EXT:COUP?
Related
Command
[:SOURce]:PM:SOURce