Instructions
RIGOL Chapter 2 Command System
2-56 DSG800 Programming Guide
[:SOURce]:PULM:TRIGger:DELay:STEP
Syntax [:SOURce]:PULM:TRIGger:DELay:STEP <value>
[:SOURce]:PULM:TRIGger:DELay:STEP?
Description Set the step of pulse trigger delay.
Query the step of pulse trigger delay.
Parameter
Name
Type
Range
Default
<value>
Real
10ns to 170s
100us
Explanation When <value> is set in "Number" form, the default unit is s. Besides, <value>
can also be set in "Number + Unit" form; for example, 50ms.
After the trigger delay step is set, you can rotate the knob to modify the trigger
delay at the current step. At this point, you can query and set the trigger delay
using the [:SOURce]:PULM:TRIGger:DELay command.
Return Format
The query returns the trigger delay step. For example, 5.000000000s.
Example :PULM:TRIG:DEL:STEP 5 /*Set the trigger delay step to 5s*/
:PULM:TRIG:DEL:STEP?
Related
Command
[:SOURce]:PULM:TRIGger:DELay
[:SOURce]:PULM:TRIGger:EXTernal:GATE:POLarity
Syntax [:SOURce]:PULM:TRIGger:EXTernal:GATE:POLarity NORMal|INVerse
[:SOURce]:PULM:TRIGger:EXTernal:GATE:POLarity?
Description Set the polarity of the external gated signal.
Query the polarity of the external gated signal.
Parameter
Name Type Range Default
NORMal|INVerse Discrete NORMal|INVerse NORMal
Explanation When the trigger mode of pulse modulation is set to "Ext Gate", the RF signal
generator receives the external gated signal input from the [TRIGGER IN]
connector at the rear panel. At this point, you can set the polarity of the external
gated signal using this command.
NORMal: set the polarity of the external gated signal to "Normal".
INVerse: set the polarity of the external gated signal to "Inverse".
When the modulation source is set to "Ext", this command is invalid.
Return Format The query returns NORMAL or INVERSE.
Example
:PULM:TRIG:EXT:GATE:POL INV
:PULM:TRIG:EXT:GATE:POL?
Related
Commands
[:SOURce]:PULM:TRIGger:MODE
[:SOURce]:PULM:SOURce