Instructions

Chapter 2 Command System RIGOL
DSG800 Programming Guide 2-57
[:SOURce]:PULM:TRIGger:EXTernal:SLOPe
Syntax [:SOURce]:PULM:TRIGger:EXTernal:SLOPe POSitive|NEGative
[:SOURce]:PULM:TRIGger:EXTernal:SLOPe?
Description Set the polarity of the effective edge of the external trigger pulse.
Query the polarity of the effective edge of the external trigger pulse.
Parameter
Name
Type
Range
Default
POSitive|NEGative Discrete POSitive|NEGative POSitive
Explanation When the trigger mode of pulse modulation is set to "Ext Trig", the RF signal
generator receives the external trigger signal input from the [TRIGGER IN]
connector at the rear panel. At this point, you can use this command to set the
trigger edge of the external trigger signal.
POSitive: set the polarity of the effective edge of the external trigger pulse to
"Pos".
NEGative: set the polarity of the effective edge of the external trigger pulse to
"Neg".
When the modulation source is set to "Ext", this command is invalid.
Return Format The query returns POSITIVE or NEGATIVE.
Example :PULM:TRIG:EXT:SLOP NEG
:PULM:TRIG:EXT:SLOP?
Related
Commands
[:SOURce]:PULM:TRIGger:MODE
[:SOURce]:PULM:SOURce