[:SOURce]:PULM:WIDTh:STEP
Syntax [:SOURce]:PULM:WIDTh:STEP <value>
[:SOURce]:PULM:WIDTh:STEP?
Description Set the step of the width of the pulse modulating signal.
Query the step of the width of the pulse modulating signal.
Parameter
Name
Type
Range
Default
<value>
Real
10ns to 10s
100us
Explanation When <value> is set in "Number" form, the default unit is s. Besides, <value>
can also be set in "Number + Unit" form; for example, 3000ms.
After the pulse width step is set, you can rotate the knob to modify the pulse
width at the current step. At this point, you can query or set the pulse width
using the [:SOURce]:PULM:WIDTh command.
Return Format The query returns the step of the width of the pulse modulating signal. For example,
3.000000000s.
Example :PULM:WIDT:STEP 3
:PULM:WIDT:STEP?
Related
Command
[:SOURce]:PULM:WIDTh