Instructions
RIGOL Chapter 2 Command System
2-74 DSG800 Programming Guide
[:SOURce]:SWEep:STEP:STOP:LEVel
Syntax [:SOURce]:SWEep:STEP:STOP:LEVel <value>
[:SOURce]:SWEep:STEP:STOP:LEVel?
Description Set the stop level of the sweep.
Query the stop level of the sweep.
Parameter
Name
Type
Range
Default
<value> Real -110dBm to 20dBm -20dBm
Explanation When <value> is set in "Number" form (for example, 2), the default unit is dBm.
When <value> is set in "Number + Unit" form (for example, 2dBm), the stop
level displayed in the interface of the RF signal generator is related to the setting
of Level Unit.
— When the level unit is "dBm", 2.00dBm is displayed.
— When the level unit is "dBmV", 48.99dBmV is displayed.
— When the level unit is "dBuV", 108.99dBuV is displayed.
— When the level unit is "Volts", 281.50mV is displayed.
— When the level unit is "Watts", 1.58mW is displayed.
The default unit of the return value is dBm.
After the stop level is set, you can rotate the knob to modify the stop level at the
current step. At this point, you can query or set the current step using the
[:SOURce]:SWEep:STEP:STOP:LEVel:STEP command.
Return Format The query returns the stop level of the sweep. For example, 2.000000.
Example
:SWE:STEP:STOP:LEV 2dBm
:SWE:STEP:STOP:LEV?
Related
Command
[:SOURce]:SWEep:STEP:STOP:LEVel:STEP